ZEUS-200S/150 ZO-OM-007
REV : 08 9 ZERONE
Check the followings before using this product.
Warning
①
Warning : To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth
②
Use the power code provided by the equipment supplier or a other test-completed
product that has equal quality.
③
Don't use a extended code that has no ground connection.
④
Connection for potential equalization : Connection pins for potential equalization cable
between RF equipment and potential equalization bar in or room. The yellow-green
potential equalization cable supplied by the manufacturer must be used.
①
Don't place any equipment on Electrosurgical Unit, which can prevent this unit
from being cooled.
②
Do not leave any water container on the main system. If any water is spilled, wipe it
up immediately, and never run the system if water is soaked into it.
③
Don't reduce the volume of display indication sound, which provides alarms to the
operator during an operation of the surgical unit.
④
When using Smoke Evacuator connecting Electrosurgical Unit, keep them apart
each other, and set up the output sound at the maximum level so that the output
sound and the alarm indication sound can be heard well.
⑤
Electrosurgical Unit should be kept apart, as much as possible, from other electronic
equipment because the use of the unit can have negative effects on other medical
units that are using other electricity.
⑥
Check the linkage condition of equipment ground.
⑦
Impressed voltage (AC120V/AC230V) should be used in accordance with the rating.
Before providing power, check an operation voltage and power frequency marked in
the label on the back of the product. Inappropriate power could cause output
fluctuation and a direct fault in the product. To prevent over-standard voltage
fluctuation, it is recommended that Automatic Voltage Regulator (AVR) be used.
3. Application Notice
Electrosurgical Unit is supposed to generate high voltage and current for its
Application. To avoid exposing a patient, an operator and the third parties
into a dangerous situation, operation should be carefully conducted, and
safety rules should be strictly kept.