PSM Training User Guide 

 

Page 5 of 31

 

Ph: 1 (443) 569-3603                                                    Fax: 1(443) 926-9402                                     Web: www.zephyranywhere.com 

9700.0142                                                             © Zephyr™ Technology 2014                                                                 2014-FEB-11                                                                                                                                        

 

1.  Introduction 

 
This document contains information required to operate the 

Zephyr™ PSM Training system incorporating 

OmniSense Application software, using Zephyr components which support the ECHO radio network type. 
 

1.1  System Description 

 
 
The PSM Training System is a remote physiological monitoring system that utilizes Zephyr

’s patented 

BioHarness™ product. The BioHarness™ allows the measurement of: 

 

Heart Rate 

 

Breathing Rate 

 

Activity level 

 

Subject orientation (posture) 

 

Estimated Core Temperature 

 

A variety of physiological parameters using automated fitness tests 

– details are in the 

OmniSense Analysis Help file

 

 

Data from the BioHarness™ is used to generate a color-coded Subject Status indication (ROG): 

 

Green   

 

Orange 

 

Red 

 

 
Data is relayed over a radio network. This configuration will support a maximum of 50 subjects. 
 
Individual subject data is updated every 2.5 seconds. 
 
Specific BioHarness 3.0 hardware is required for a PSM Training system. This hardware is dual 
Bluetooth/ECHO.  
 
The ECHO system does not support Bluetooth-only BioHarness modules.  
 
 
 

 

Содержание PSM Training ECHO

Страница 1: ...PSM Training ECHO User Guide...

Страница 2: ...me of publication but is subject to change without notice Zephyr Technology assumes no responsibility for errors or omissions and disclaims responsibility for any consequences resulting from the use o...

Страница 3: ...2 4 Operational Startup 15 2 5 Live Screen Operation Checks 16 2 6 Field Repeater Units 17 2 7 GPS Support 18 2 7 1 Configuring a BioHarness for use with a GPS 19 3 LOGGING 22 3 1 Download Device Log...

Страница 4: ...am ECHO 2 4 GHz 802 15 4 Radio Network type used by this system Gateway Receiver device connected to PC HR Heart Rate HR max Maximum heart rate of an individual subject LED Light Emitting Diode KML Ke...

Страница 5: ...product The BioHarness allows the measurement of Heart Rate Breathing Rate Activity level Subject orientation posture Estimated Core Temperature A variety of physiological parameters using automated...

Страница 6: ...ystem Diagrams 1 BioHarness TM chest strap up to 50 2 BioHarness TM 3 0 ECHO module up to 50 3 ECHO gateway additional repeaters can be used in the field to extend the network 4 USB connector 5 PC pla...

Страница 7: ...tion to PC and power outlet simultaneously for use Single Device Charge Cradle Mini USB connectivity for device configuration log download Connect to USB wall car charger for charge only ZEPHYR ECHO G...

Страница 8: ...Fax 1 443 926 9402 Web www zephyranywhere com 9700 0142 Zephyr Technology 2014 2014 FEB 11 Field Repeater Integrated rechargeable battery power supply 24 hr battery life 5200mAh 3 hr charging Increas...

Страница 9: ...here com 9700 0142 Zephyr Technology 2014 2014 FEB 11 1 4 Software Installation Follow the instructions in the OmniSense Software Installation Guide to Install OmniSense Live Analysis modules includin...

Страница 10: ...ork Setup Before adding any hardware components to the system database radio network type must be set up or confirmed in the OmniSense Live module The Radio Network Type must be set to ECHO 2 2 Databa...

Страница 11: ...are allocation in OmniSense Setup Hardware tab 2 Check components match those that are assigned to the subject 24 3 Power on components check LEDs To power OFF BioModule detach from strap press hold c...

Страница 12: ...at rear should be flush with strap shown un tensioned here when subject inhales and chest is fully expanded 2 Centre line of the device should be directly under the armpit for optimum breathing detect...

Страница 13: ...oth or Logging error then a hardware fault is likely replace the device in the interim Bluetooth ECHO Logging Battery HR Detect Downloading DEVICE STATE IN CRADLE Connected Error Always off Charging C...

Страница 14: ...e orientation of the device varies and accelerometer data must be interpreted correctly in order to display posture data There are three variations of garment type The garment device type can be selec...

Страница 15: ...02 Web www zephyranywhere com 9700 0142 Zephyr Technology 2014 2014 FEB 11 1 Set up at preferred location ACTIVITY AREA See also suggestions on optional field repeater units 2 Connect Gateway to PC 4...

Страница 16: ...CE POWER ON Heart Rate 5 seconds Breathing Rate 30 45 seconds Core Temperature 1 second Activity Posture 1 second 3 WAIT 1 2 minutes It takes the Gateway 1 2 minutes for all network connections to est...

Страница 17: ...xtend absolute range of the system or provide better coverage of an area Field repeaters transmit directly to the base ECHO gateway but NOT to each other They cannot be used to extend the system range...

Страница 18: ...nfigured using the Zephyr Cfg Tool see the next section LIVE MODULE No GPS data can currently be displayed in the Live module This function is supported by PSM Responder and some PSM Defense systems o...

Страница 19: ...r a kml file can be exported from a session containing GPS data for import into Google Earth Streaming GPS data cannot be displayed on a BioGauge in OmniSense live using ECHO this can only be done whe...

Страница 20: ...2 Zephyr Technology 2014 2014 FEB 11 2 Windows will scan for any Bluetooth devices It may take several attempts When the Qstarz has been detected right click on the icon and select Properties from the...

Страница 21: ...r the GPS Mac address in the GPS Device field and click the Set BT Dev 9 button 5 In the User Cfg tab confirm that the Log Format is set to Summary the default or Summary and Waveform 6 In operation s...

Страница 22: ...se If the BioHarness module is configured appropriately ECG and accelerometer waveform csv files amongst others can be generated 3 1 Download Device Logs 3 1 1 Download to OmniSense Database A button...

Страница 23: ...t section Avoid storing in an uncharged state for prolonged periods A top up charge of the battery is recommended once a month if the devices are not used regularly to maintain optimum battery conditi...

Страница 24: ...se shallow charge discharge cycles prolong battery life Avoid storing the devices fully discharged for prolonged periods it may damage reduce battery life A top up charge is recommended once per month...

Страница 25: ...o Firmly attach the Velcro fastenings together and do not wash with other garments which may be damaged by these fastenings o Use a washing pouch if possible o Use soap or mild detergent but NOT steri...

Страница 26: ...used to control settings Care should be taken when changing any parameters in the device Record any settings before you change them and seek advice from Zephyr if you are in doubt about a setting 1 C...

Страница 27: ...ed to the device You must use the utility ZUSBUpdater exe in conjunction with a suitable img firmware image file Hardware compatibility is checked so you should not be able to update a device with fir...

Страница 28: ...ty 80 after 300 deep discharge cycles 4500hrs of use Storage Life 6 months between charges for optimal recharge monthly Strap Sizes Small XS M 27 38 inches Large M XL 33 43 inches Wash durability Typi...

Страница 29: ...nd logs data internally and can record up to 24 hours of data in a single session All logged data is time stamped using an internal clock in the BioHarness Module The internal clock is not used to tim...

Страница 30: ...device at apex of rib curvature 8 3 General Issues Parameter Indication Possible Cause Check Remedy Communications No data in BioGauges Radio network error Exit OmniSense disconnect the Gateway recei...

Страница 31: ...This warranty shall also be voidable by zephyr or its licensed distributors If 1 Zephyr reasonably believes that the PSM TRAINING system has been used in a manner that would violate the terms and cond...