Operation of the device
000000-1272-175-GA-GB-060613
64
Treatment STANDBY mode
After the system has been switched on and the internal system test has been
completed, the system enters this mode automatically unless the instrument
is switched to Diagnosis mode by the user.
Fig. 26
Treatment STANDBY mode
All treatment-related laser parameters are displayed on the control panel and
can be changed in this mode.
The following laser parameters can be set in this mode: energy, pulse mode
and aiming beam intensity.
Each time the system is restarted, the energy and the aiming beam intensity
values most recently selected will be used.
•
To set individual laser parameters, touch the appropriate parameter
button and set the value required by turning the control knob.
•
Select the desired pulse mode (single, double or triple pulse) by touching
the appropriate button. The reverse-colored button (blue on a white
background) is activated.
•
Pressing the control knob takes you directly to the entry field for the laser
energy. Rotate the knob to set the energy required.