manualshive.com logo in svg
background image

 

SET-UP 

Transillumination top 450 mot. 

Mounting the … 

Carl Zeiss 

 
 

 

08/2012 

435500-9000-701 

13 

 

Switch off the Axio Zoom.V16 microscope and illumination. 

 

Remove the base insert plate (Fig. 2/

1

). To do this loosen the short socket head screws (Fig. 2/

2

), lift 

the insert plate (use recessed grip (Fig. 2/

4

) on the right edge of stand base for the purpose) and lift 

securely with both hands. 

 

Loosen the three socket head screws (SW 3) in the adapter ring (Fig. 2/

3

) using a ball-headed 

screwdriver and remove the adapter ring. 

 

Mount the transillumination top 450 mot. (Fig. 2/

6

) on the base plate (Fig. 2/

5

).  

 

 

Fig. 3 

Mounting the transillumination top 450 mot. 

 

 

For parallel alignment push the transillumination top slightly backwards with the finger (Fig. 3/

8

) and 

screw it hand-tight to the base plate using the four countersunk captive hexagon socket screws (SW 3) 
(Fig. 3/

7

). 

 

The transillumination top may not be connected to the mains supply until the shipping screws 
(Fig. 3/

9

) have been removed; damage may otherwise be caused by the initialization runs thus 

triggered.  

 

The two shipping screws (Fig. 3/

9

) must be removed and kept in a safe place. 

 

Insert the magnet-coded receptacle d = 84 mm (Fig. 3/

6

) with polarizer (Fig. 3/

5

) only for polarization 

applications (pay attention to markings for north-south orientation). Attach the respective analyzer to 

the objective.  

 

If the magnet-coded receptacle d = 84 mm (Fig. 3/

6

) is mounted or removed at a later time, the 

microscope system must be switched on again/restarted for the new status to be recognized.  

 

Place the insert plate (Fig. 3/

4

) on the transillumination top. 

 

Insert assembly aid d = 120 mm (Fig.  3/

3

) and align with the threaded holes. This will prevent the 

screws and small parts from falling into the transillumination top. 

 

Screw the insert plate onto the transillumination top using the three hexagon socket screws (Fig. 3/

1

).  

 

Remove assembly aid d = 120 mm. Insert the glass plate 120 mm (Fig. 3/

2

). 

 

To prevent eye injury due to scattered light, attach and swing in the antiglare shield (Fig. 1/

I

) on 

the right or left of the mount or revolving nosepiece. 

Содержание Transillumination top 450 mot.

Страница 1: ...Operating Manual Transillumination top 450 mot for Axio Zoom V16 ...

Страница 2: ......

Страница 3: ...ed a high quality product from Zeiss Before using the instrument the first time please read this operating manual in order to maintain the high quality of the instrument and ensure reliable work with it for a long time ...

Страница 4: ...subject to updating or revision Unless explicitly authorized dissemination or duplication of this document and commercial exploitation or communication of its contents are not permitted Persons in contravention of this copyright shall be liable to pay compensation for damages All rights reserved in the event of a patent grant or registration of a utility patent Published by Carl Zeiss Microscopy G...

Страница 5: ...ication for use 9 2 2 Package check list with accessories 9 2 3 Microscope system with transillumination top 450 mot 10 2 4 Technical data 11 3 SET UP 12 3 1 General information 12 3 2 Mounting the transillumination top 450 mot 12 3 3 Connecting the transillumination top 450 mot 14 4 OPERATION 15 4 1 Control elements 15 4 2 Switching on 17 4 3 Switching off 17 4 4 Troubleshooting 18 4 5 Functional...

Страница 6: ... WEEE Directive 2002 96 EC It carries the mark The operating manual contains all information and warnings the operator must comply with The following warning and information symbols are used in this operating manual CAUTION This symbol indicates a potential hazard to the user CAUTION Optical radiation is emitted Do not look into the laser beam This may be harmful to the eyes see Section 2 3 CAUTIO...

Страница 7: ...ty claims invalid The device may only be operated by instructed personnel The instruments may only be operated by trained personnel who are aware of the possible dangers involved with microscopy and the application concerned The microscope is a high precision instrument that can be impaired in its performance or destroyed when handled improperly If it is determined that protective measures are no ...

Страница 8: ...irectly into the light beam This can cause damage to the eyes and skin The transillumination top 450 mot generates extremely high energy light in the visible and invisible range The related instructions for safety and use must be strictly adhered to 1 3 Notes on warranty The manufacturer guarantees that the instrument is free of material or manufacturing defects when delivered Possible defects mus...

Страница 9: ...d in 435500 9000 000 Can be screwed onto the carrier nosepiece from the left or right Magnet coded receptacle d 84 mm included in 435500 9000 000 For 84 mm accessories Clear glass plate d 120 mm included in 435500 9000 000 Scratch proof with strongly reduced autofluorescence Assembly aid d 120 mm included in 435500 9000 000 For covering the transmitted light aperture during assembly Power supply 2...

Страница 10: ...ate tube Fluar Illuminator Z mot G Camera adapter with microscope camera AxioCam H Reflected light illumination slit ring illuminator with cold light source CL 6000 LED I UV visual glare shield K Stand focus motor with 490 mm profile column and MaRC control unit on base plate 450 L Transmitted light illuminator transillumination top 450 mot M Specimen stage gliding stage N SYCOP 3 system control p...

Страница 11: ...ble relative humidity max 75 Air pressure800 hPa to 1060 hPa Degree of pollution2 Optical risk group classification acc to DIN EN 62471 2009 Axio Zoom V16 microscope system with Transillumination top 450 mot LED risk group 1 acc to DIN EN 62471 2009 Radiation apertures Axio Zoom V16 microscope system with Transillumination top 450 mot conically upwards from the opening Table power supply Operation...

Страница 12: ...facturer or dispose of it properly The following system requirements must be observed Use of stand 450 from 435430 9902 000 or later models Use of focus motor 2 or 3 as CAN distributor Optional software EMS 3 435610 9010 000 or later model ZEN 2011 Blue Edition or later model Optional mechanical stages Mechanical stage S 150x100 R 435465 9010 000 or later models Mechanical stage S 150x100 mot CAN ...

Страница 13: ...ping screws Fig 3 9 have been removed damage may otherwise be caused by the initialization runs thus triggered The two shipping screws Fig 3 9 must be removed and kept in a safe place Insert the magnet coded receptacle d 84 mm Fig 3 6 with polarizer Fig 3 5 only for polarization applications pay attention to markings for north south orientation Attach the respective analyzer to the objective If th...

Страница 14: ... A BNC cable can be connected to the Trigger Fig 4 2 plug in point to transmit a trigger signal Alternatively the foot switch S can be used to switch the illumination of the transillumination top on off via the Foot switch Fig 4 1 plug in point Connecting to the mains supply Connect the power supply unit to the 24V DC Fig 4 4 plug in point Plug the power cable of the PSU into a mains socket Upon c...

Страница 15: ... to the microscope system Mode keypad Fig 5 2 Upon switching on the trans illumination top the Best Mode function is always activated LED lights up when the respective function is activated by pressing a button BF button Brightfield illumination BF button Brightfield illumination with additional contrast enhancement RC button Relief contrast by oblique illumination DF button One sided darkfield Li...

Страница 16: ...conds the parameters set for zoom position magnification illumination method BF BF RC DF brightness and respective adjustment function of the configured system are saved Briefly pressing the respective button resets the saved parameters Saved settings can no longer be retrieved after changing the objective Pressing the Memory button has no effect the button s LED flashes If an objective for which ...

Страница 17: ...ions during the focusing movement Switch on the microscope system via HIP or on the SYCOP Switch the transillumination top by pressing the on off button Fig 4 5 on the rear side Select the desired magnification on the microscope and adjust the focus as necessary Select the illumination method for transmitted light on the Mode keypad Open Shutter as necessary If a Foot switch S is connected it can ...

Страница 18: ...ert plate missing Attach the insert plate see Section 3 2 Unusual noises during initialization of the transillumination top Shipping braces have not been removed Disconnect the transillumination top from the mains and remove the shipping braces see Section 3 2 Collision noises of the apertures of the transillumination top Use of an insert plate that is not system compatible Replace the insert plat...

Отзывы: