Safety measures
S100 / OPMI pico
Version 9.0
Page 16
G-30-1781-en
•
If the xenon lamp is used beyond its maximum service life of 500 hours
(see chapter "Maintenance intervals" on Page 222), sudden failure may
occur and interrupt operation. Change to the backup lamp or replace the
lamp module in a timely manner.
•
If the halogen lamp is used beyond its maximum service life of 50 hours,
sudden failure may occur and interrupt operation. Replace the halogen
lamp in a timely manner (see chapter "Service and maintenance" on
Page 224).
•
Do not pull on the power cord or any other connecting cables.
•
If there is a failure and you cannot get any help from the chapter "What
to do in the event of malfunctions", attach a sign "Out of order" to the
system and contact ZEISS Service representative.
•
Constantly monitor the system during use.
•
Always switch off the system every time before:
- connecting it to / disconnecting it from line power
- if it will not be used for a prolonged period of time
- or if you want to clean its surface.
•
If the light guide is not connected at both ends, there may be a risk of fire
or burning when the light source is activated.
•
Risk of crushing - mind your fingers!
Fingers may be crushed in the areas marked with the "Risk of crushing"
sign.
Do not touch these areas while the device is being moved.
After every use
•
Always use the master switch to turn off the device if it is not in use.
•
Insufficient, incorrect or wrong cleaning or disinfection contrary to the
instructions of the Instructions for Use may expose the patient or medical
staff to a considerable risk of infection. For more information, please see
"Maintenance of the device" in the chapter "Care and maintenance".
Содержание S100
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Страница 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Страница 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Страница 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Страница 111: ...Version 9 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 58 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Страница 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Страница 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Страница 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Страница 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Страница 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...