6 Before every use
OPMI LUMERA 700
6.4 Preparing the system for sterile use
G-30-1673-en - 14.0 - 2018-07-12
127 / 236
6.4 Preparing the system for sterile use
6.4.1 Attaching sterile accessories
CAUTION!
Risk of infection caused by missing sterile accessories!
The patient or user can become infected if sterile accessories are
not used.
u
Use the device only with suitable sterile accessories.
u
Disinfect, clean and sterilize sterile accessories before each use.
This also applies to the first use after delivery. These procedures
may only be performed by properly trained personnel.
Material
• Sterilizable cap
• Sterilizable cap, 22 mm
• Sterilizable cap, 49 mm
Documents
• "Preparation of resterilizable products"
manual
Prerequisite
þ
The resterilizable caps have been sterilized.
Action
1. Slide the sterilizable caps onto the handgrips on the surgical
microscope.
2. Turn the sterilizable caps in such a way that the elevations for
the buttons precisely fit on the handgrip buttons.
3. Slide the 22 mm sterilizable cap onto the adjustment knob on
the surgical and assistant's microscopes.
Содержание opmi lumera 700
Страница 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Страница 10: ...Empty page for your notes...
Страница 14: ...Empty page for your notes...
Страница 30: ...Empty page for your notes...
Страница 130: ...Empty page for your notes...
Страница 200: ...Empty page for your notes...
Страница 222: ...Empty page for your notes...
Страница 226: ...Empty page for your notes...
Страница 230: ...Empty page for your notes...
Страница 235: ...Empty page for your notes...