MAINTENANCE, FUSE CHANGE, AND SERVICE
MCU 2008
Maintenance
Carl Zeiss
M60-2-0019 e 05/2012
15
5
MAINTENANCE, FUSE CHANGE, AND SERVICE
5.1
Maintenance
Before cleaning the MCU 2008 it is imperative to disconnect the power plug.
•
The MCU 2008 must be cleaned using a slightly moistened, not dripping cloth.
No moisture must get into the instrument.
5.2
Fuse Change
Before changing the fuses it is
imperative to disconnect the power
plug.
The fuse holder is located on the back of the
MCU 2008.
In the event of a fuse failure, first the cause must
be identified and the technical defect, if any, must
be eliminated.
•
Disconnect the power plug.
•
Pull the fuse holder (Fig. 5-1/
1
) frontward out of
the fuse compartment.
•
Replace the defective fuse
T 2.0 A/H
.
•
Push the fuse holder back into the fuse
compartment until it snaps into place.
•
Reconnect the power plug.
Fig. 5-1
Replacing fuses