NOTES ON DEVICE SAFETY
LSM 510
DuoScan
Carl Zeiss
Notes on Setting up the Microscope System
LSM 510 META
DuoScan
1-12
B 45-0021 e
03/06
1.4
Notes on Setting up the Microscope System
Installation and commissioning of the LSM 510 and LSM 510 META system must be performed
by authorized Carl Zeiss service staff. The system should not be used prior to instruction by a
Carl Zeiss representative.
The LSM 510 or LSM 510 META laser scanning microscope is delivered in several crates.
The LSM 510 and LSM 510 META must be set up so as to ensure that the minimum clearance
between the wall and the rear of the system is no less than 0.5 m. This clearance is needed for
adjustment and maintenance operations.
Do not set up the unit in the proximity of heat sources such as radiators or direct sunlight. To avoid heat
build-ups, the ventilation slots on the microscope system must not be covered up.
The system must not be set up in areas with potential danger by explosives.
The unit must be connected to a properly installed socket outlet with earthing contact by means of the
mains cables supplied. Continuity of PE connection must not be affected by the use of extension leads.
The system contains components with dangerous voltage. The system must not be opened by
anybody else than authorized Carl Zeiss Service staff. Before opening the main plug has to be
disconnected.
Before connecting the mains cables, please check whether your mains voltage corresponds to
the voltage specified on the rating plate of the laser module.
For reasons of laser safety, all ports must either be equipped with the corresponding device
(scan head, camera, HBO lamp etc.) or covered with the counterpart of the laser safety kit
provided.
Maintenance, repair, modification, removal or exchange of components, or other interference
with the equipment beyond the operations described in this manual may only be carried out by
the manufacturer Carl Zeiss or by persons expressly authorized by Carl Zeiss to do so.
This applies especially to the microscope system, the laser scanning module, lasers, the PC
system, the power supply units, cable connections and other system components.
Please note that the LSM
510 and LSM
510
META are high-precision opto-electronic
instruments. Inexpert handling may easily impair their function or even damage them.
The openings for ventilation must not be covered.
There are hot surfaces on the HBO and HAL lamp.
When sliding the compact Lasermodule V in and out of the System electronic rack take care not
to catch your fingers.
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......