Removing and Inserting the Illumination Optics Unit
The illumination optics unit (Fig. 35/
2
) can easily be removed and inserted after the removal of the sample
3.4 Removing and Inserting the Sample Chamber
To protect the front lenses, secure the appropriate protective caps (Fig. 35/
3
) over each of the
illumination optics (Fig. 35/
2
) and protective cap (Fig. 35/
4
) over the detection optic (Fig. 35/
5
). The
illumination optics can then be unscrewed by turning anticlockwise.
To insert the illumination optics, proceed in the reverse order by screwing it clockwise into the unit until
finger-tight.
Finally, the protective caps can be removed and the sample chamber inserted.
After the illumination optic has been exchanged, the currently used optic must be registered in the ZEN
software. The necessary software interface can be found in the Objectives tool window under the
Maintain tab.
1
Front system cavity door
4
Protective cap for detection optics unit
2
Illumination optics (2×)
5
Detection optics unit
3
Protective cap for illumination optics unit (2×)
6
Threaded shaft for illumination optics unit (r and l)
Fig. 35
Removing and inserting the illumination optics