Содержание EM 10 A

Страница 1: ......

Страница 2: ...A Each of these four lenses can be displaced against a restoring spring as a whole centering With C2 and PI it is possible in addition to displace one polepiece with respect ot the other this tilting...

Страница 3: ......

Страница 4: ...in and use specimen drives to move image detail to central position between the two image positions Repeat polarity reversal until image detail appears in exactly identical position with both polariti...

Страница 5: ......

Страница 6: ...is not satisfactory or after disassembling PI C 1 35 Set magnification to 20 000 and note direction of image sweep Switch back to 6 300 C 1 36 Use PI tilting screws 1 0 3 to displace image detail in...

Страница 7: ......

Страница 8: ...and to adjust for m a x i m brightness turning each beam alignment potentiometer back and forth results in equal loss of brightness in both ways if aperture image is displaced during this procedure re...

Страница 9: ......

Страница 10: ...C 1 46 Turn C2 switch clockwise from minimum spot setting C2 over excited and note sweep direction of growing spot Reset C2 switch for minimum spot size C 1 47 Use C2 tilting screws 1 0 1 1 to displa...

Страница 11: ......

Страница 12: ...cation selector to 1 000x Repeat C 1 53 and C 1 54 until obtaining minirmun spot size Set magnification selector to 6 300 Center bright spot acc to C 1 54 C 1 56 Repeat final check as in C 1 45 If nec...

Страница 13: ......

Страница 14: ...ce pole plate now displaceable Close column and pump down without inserting cooling plate C 1 62 Further preparatory steps Switch off all lenses keys on left alignment panel Fully withdraw all apertur...

Страница 15: ......

Страница 16: ...in focusing position of Coarse knob turn 5 steps and note current center center point of image rotation C 1 644 Use alignment screws 10 7 to displace current center image detail coinciding with curren...

Страница 17: ......

Страница 18: ...up t o 1 5 mm 114 length of center mark off screen center i n any direction C 1 66 Precision alignment on current center normal polarity C 1 661 Turn Coarse knob 5 21 3 steps from focus and note new c...

Страница 19: ......

Страница 20: ...by magnetically a t t r a c t e d t o o l Place top cooling p l a t e 3 on copper studs 5 and f i x with 3 brass screws cf p a r t D Turn alignment screws 10 7 anti clockwise p u l l out u n t i l mee...

Страница 21: ......

Страница 22: ...rn switch 1 25 clockwise until shadow image of new aperture is somewhat smaller than large screen C 2 14 Using axial and transversal aperture drives with left hand center aperture image on large scree...

Страница 23: ......

Страница 24: ...cus c e n t r a l spot of d i f f r a c t i o n p a t t e r n with potentiometer 5 22 I f necessary correct astigmatism of c e n t r a l spot with PI stigmator cf p a r t B C 2 25 I n s e r t objectiv...

Страница 25: ......

Страница 26: ...een edge with left hand C 2 34 Checking alignment of intermediate image aperture Set magnification to 6 300 aperture image decreases centeredly to screen center Otherwise recheck alignment of P1 and P...

Страница 27: ......

Страница 28: ...ion signals burnt out filament Make sure that beam tilt system is switched off key 4 15 C 3 22 Set magnification to 1 000 and switch off P2 key 3 16 C 3 23 Press all three keys I 11 I11 of aligment ai...

Страница 29: ......

Страница 30: ...i directional blur of image detail appears C 4 4 Use right potentiometers 3 23 alternately to minimize blur use small screwdriver complete elimination of blur impossible C 4 5 Release key 3 19 and pre...