SAFETY INFORMATION
Lattice Lightsheet 7
Power Requirements
ZEISS
04/2021 V_01
000000-2324-960
13
Power supply unit (PSU) and laser module (LM) unit can be housed in the component rack (see Fig. 3). The
components 6-point power strip is connected to the multipoint connector for components and is attached
to the rack.
Fig. 3
Component rack with power supply unit (PSU), laser module (LM), and motion controller (MC)
for Lattice Lightsheet 7 (rear panels)
The PSU and the LM are pre-mounted, the MC will be mounted on-site.
The mains plug may only be inserted into a socket with a protective contact. The protective effect must
not be overridden by an extension cable without a protective conductor.
Detachable power cords should not be replaced with inadequately assessed power cords. Only the
specified power cords may be used.
Only the connection cables supplied with the device may be used. The use of commercially available
cables for use with the LLS7 is prohibited.
A safe disconnection from the power supply network is only guaranteed by pulling the power plug.
The switch on the controller only switches to standby mode.
Danger of voltage from contact with live parts. This can lead to burns and a fatal electric shock.
Switch off the microscope before cleaning. Do not touch any live parts. Disconnect live parts from
the power supply.