3.4
Setting the microscope in brightfield
Place specimen on the sample stage with the cover slip pointing upwards.
The condenser 0.9 AS (3-2/
3
) is aligned with
the specimen stage and firmly attached to it. It
contains an aperture diaphragm to control the
image contrast. The condenser lever (3-2/
1
) is
attached to the bottom of the condenser via a
knurled knob (3-2/
4
).
Swing in the 10x objective. When using an
objective with magnification 10
or higher,
switch the condenser lever (3-2/
1
) to free
passage by turning the lever (3-2/
2
) to the right
until stop.
The condenser lever (3-2/
1
) can be switched in two stop positions:
Lever (3-2/
2
) to the right:
free passage (to be used for objectives
10
).
Lever (3-2/
2
) to the left:
additional lens swung in (to be used for objectives < 10
).
The aperture diaphragm can also be set to both stop positions via the lever (3-2/
2
).
Turning to the right within the stop position:
aperture diaphragm is opened
Turning to the left within the stop position:
aperture diaphragm is closed.
Fig. 3-2
Setting the condenser lever
Содержание KF 2 ICS
Страница 1: ...KF 2 ICS Transmitted Light Microscope Operating Manual...
Страница 6: ...Carl Zeiss KF 2 ICS VI B 40 028 e 10 97 Overall view of the KF 2 ICS microscope...
Страница 8: ...Carl Zeiss KF 2 ICS 1 2 B 40 028 e 10 97...
Страница 10: ...Carl Zeiss KF 2 ICS 1 4 B 40 028 e 10 97 1 3 Modules Fig 1 1 KF 2 ICS microscope equipment...
Страница 14: ...Carl Zeiss KF 2 ICS 2 2 B 40 028 e 10 97...
Страница 20: ...Carl Zeiss KF 2 ICS 3 2 B 40 028 e 10 97...
Страница 32: ...Carl Zeiss KF 2 ICS 4 2 B 40 028 e 10 97...
Страница 38: ...Carl Zeiss KF 2 ICS A 2 B 40 028 e 10 97...
Страница 40: ...Carl Zeiss KF 2 ICS A 4 B 40 028 e 10 97...