NOTICE
You probably cannot calibrate successfully with a dirty or damaged cal‐
ibration panel.
The GOM software might not recognize dirty or damaged calibration pan‐
els.
▶
Do not soil or damage particularly the area around the larger calibra‐
tion points in the middle!
NOTICE
When you use a wrong calibration object, the calibration is incorrect.
When the calibration object does not match the calibration object data in
the software, the measuring system provides inaccurate data. The soft‐
ware only checks the correct type of the calibration panel but not the
serial number.
▶
If you use several sensors, check the serial number of the calibration
object!
▶
Compare the serial number of the calibration object with the serial
number of the calibration object in the GOM software!
8.4.2
Prepare the Calibration
Procedure:
1. Open the calibration dialog in workspace
Set Up
with
Cal‐
ibrate Sensor
►
.
2. Enter the current measurement temperature in the dialog.
3. Confirm the prompt with
OK
.
The software resets the settings to the recommended calibration set‐
tings.
Fig. 7: Calibration of GOM Scan 1
Calibrate the Sensor
0000002121_003_EN_01-09-2021
Page 18 (24)