Preparing the
CT LUCIA 601P / 601PY and CT LUCIA 201P
Implantation
Press the injector plunger forward and then continuously apply and release pressure. Continue this “pulsing action”, to ensure
the IOL advances smoothly. The IOL will slide smoothly into the eye.
1
We recommend to use ZEISS OVDs and OVD should be at room temperature.
2
For lenses having less than 8 diopters of optical power, it is necessary to push down on the optic body with a forceps or OVD
cannula, during wing closure to ensure that the lens folds in the correct direction. When the “click-lock” mechanism is engaged,
the lens is securely in place.
To remove the lens positioning clip, place two fingers at the top
of the lens holder and use your thumb underneath the clip to
gently release it. Then lift the clip straight up.
Cover the whole lens with a
generous
amount
of OVD
1
. In addition,
apply a small amount of OVD
1
onto the silicon tip of the plunger.
Avoid touching the lens.
Check that the lens is in the correct position, central and secure in
the loading chamber.
Close the lid of the loading chamber
2
and listen for the click, to ensure
the “click-lock” mechanism is engaged.
To advance the lens into the injector tip, gently press the plunger forward
and push the lens into the conical tip of the injector, ensuring the lens
moves smoothly forward.
Retract the plunger slightly
to separate
the silicone cushion tip from the IOL and ensure the haptics are properly
positioned. The IOL is now ready for implantation.
IMPORTANT:
The lens should be injected straight away. Viscoelastic materials may lose
their lubricity if allowed to stand too long while exposed to air.