Page
10
Installation Instructions
a
c
850
min. 800
min. 800
203
283
1048
T-T
t
2403
b
CONTURA
G2
2653
3003
c
d
e
f
g
4.2
Installation Dimensions for CONTURA G2 with opt. U-profile
(all dimensions in mm, scale 1:30)
CONTURA G2 with optional U-profile
X/Y/Z
Measuring range
Net weight of
CMM
Max. permissible
workpiece weight
1000/1200/600
ca. 2583 kg
1150 kg
1000/1600/600
ca. 3163 kg
1500 kg
1000/2100/600
ca. 5755 kg
1814 kg
c
= On-site air supply connection, precleaned compressed air
supply with pressure of 6-10 bar.
d
= Power supply (two outlets with max. 16 A) to be installed
by the customer.
e
= Telephone connection (ISDN)
f
= CMM compressed air supply connection
g
= CMM
↔
control connecting cable
t
=
Transport height without base
CONTURA G2 with optional U-profile
X/Y/Z
Measuring range
Length
- a -
Length
- b -
Length
- c -
Length
- t-
1000/1200/600
1743
2030
1200
1806
1000/1600/600
1743
2430
1000/2100/600
1743
2945
1959
Height of
secured CMM
assembly without
pallet and
without Z tower
Min. overhead clearance: 200 mm
Controller
Length: ca. 756
Width: ca. 575
Height: ca. 700
T-T = distance betw. table
and lower edge of probe
VAST
XT
probe
919
RDS probe
868
Height of
secured
CMM assy
incl. pallet
CONTURA G2
X/Y/Z
Measuring range
Max. cable leng.
CMM
↔
control
g
1000/1200/600
ca. 6300 mm
1000/1600/600
ca. 6000 mm
1000/2100/600
ca. 5700 mm