Installation Instructions
Page
11
4.3
Installation of Floor Stripe or Chain
The dangers posed by the CMM are impossible - or at
least difficult - for untrained personnel to recognize. For
this reason, a restricted-access zone must be set up
around the machine to prevent damage or injury caused
by untrained personnel.
The entrance to this area should be be clearly marked by
a warning sign reading:
"Admittance only for authorized personnel"
and may be entered only by persons who have received
proper training (operators).
min. 1000
min. 1000
min. 1000
min. 600
min. 200
min. 700
min. 100
min. 100
min. 200
Floor marking or chain
(black-and-yellow)
d
Controller
c
c
The back door of the controller cabinet must
be located at least 200mm away from the wall.
Moreover, the air exchange with the surround-
ing environment should not be blocked. The control cab-
inet should be placed so that the rear side of the
controller always faces away from the CMM.
d
CMM table (optional) with computer and printer
Note!