To ensure long life of the stage, the scuff from the specimen holders should be removed at
regular intervals. Make sure that the scuff does not get into the guiding components for the x-
movement.
(6)
Changing the specimen holder
•
Loosen two countersunk screws (2-38/
1
) using
).
•
2
) from the
4
) in an upward
direction.
•
Attach required specimen holder (2-38/
mechanical stage and tighten two countersunk
screws.
2.2.10
Changing the binocular tube
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-39/
3
) and remove available
tube in an upward direction.
•
Place dust cap (2-39/
2
) for tube lens protection
above the dovetail of the binocular tube.
•
Remove dust cap from the required tube.
•
Insert dovetail of tube (2-39/
1
) in the stand
) and align the tube.
•
3
).
Fig. 2-38
Changing the specimen holder
Fig. 2-39
Changing the binocular tube