START-UP
Axioscope 5/7/Vario
Mounting standard components
ZEISS
03/2018
430035-7344-001
45
Inserting the PL 10x/23 GW focusable Pol
eyepieces (444038-9900-000) with reticle into
the binocular photo tube with upright image
(425520-9100-000)
The PL 10x/23 GW foc. Pol eyepiece contains a
reticle firmly glued into it (cannot be changed),
which is of defined orientation. When changing
the interpu-pillary distance on the binocular photo
tube, the two eyepiece tubes follow this rotary
motion synchronously, so that the position of the
orientation grooves in the eyepiece tubes remains
unchanged.
•
Insert the eyepiece into the binocular photo
tube, fitting the locating screw (Fig. 3-5/
2
) into
the tube’s orientation groove (Fig. 3-5/
1
).
The PL 10x/23 GW foc. Pol eyepiece (444038-9900-000) can be combined with a PL 10x/23 GW foc.
eyepiece (444036-9000-000).
Inserting the eyepiece reticle
Eyepiece reticles (Fig. 3-6/
5
) can be used with
eyepieces (Fig. 3-6/
3
) marked with a red dot
R
).
•
Unscrew mounting stop (Fig. 3-6/
6
) from the
eyepiece (Fig. 3-6/
3
).
•
Pull locking ring (Fig. 3-6/
4
) out.
•
Insert eyepiece reticle (Fig. 3-6/
5
) into the
mounting stop and fix it with the locking ring.
•
Screw mounting stop (Fig. 3-6/
6
) into the
eyepiece (Fig. 3-6/
3
).
Make sure that the side of the reticle (Fig. 3-6/
5
)
with the line pattern on it is facing the mounting
stop in the tube (Fig. 3-6/
6
) (the line pattern occurs
mirror-inverted from the observer’s point of view
before the insertion). After insertion into the beam
path of the microscope it is readable true to side.
The slight image shift caused by the additional path through glass is taken into account on the diopter
scale by the fact that the zero point position is indicated not by the white dot (Fig. 3-6/
W
), but the red
dot (Fig. 3-6/
R
).
NOTE
The eyepiece reticles must be inserted under dust-free conditions. This should be carried out
only by ZEISS Service.
Fig. 3-5
Inserting the E-PL 10x/23 GW foc.
Pol eyepieces into the binocular
photo tube with upright image
Fig. 3-6
Inserting the eyepiece reticle