OPERATION
Carl Zeiss
Settings
Axio Zoom.V16
56
435080-9030-701
05/2012
Adjusting transillumination base 300
The transillumination base is connected.
•
Switch on using the rotary knob (Fig. 47/
1
) and
regulate illumination intensity. The blue power
indicator lights up when switched on.
•
Select the desired illumination method using the
setting wheel (Fig. 47/
3
).
Different effects can be achieved with the shutter slider (Fig. 47/
2
) in the individual illumination methods.
Symbol
Technical description
Applicative description
BF
Bright field
Soft contrast by oblique illumination
BF+
Higher contrast brightfield
Stronger contrast by oblique illumination
DF
All-side darkfield
Step-by-step reduction of all-side to one-side
darkfield illumination
The diaphragm slide (Fig. 47/
2
) should initially be at the rear stop in order to achieve a large
homogeneous luminous field.
Pull it slowly to the front while observing the microscopic image to achieve the desired effect.
The transmitted light illumination can be completely covered with the shutter slider (front stop)
to achieve a dark background, e.g. for fluorescence illumination.
Fig. 47 Adjusting transillumination base 300