OPERATION
Carl Zeiss
Illumination and Contrast Techniques in Transmitted Light
Axio Vert.A1
84
431030-7044-001
05/2012
•
If necessary, use the two centering keys to
precenter the iHMC module in the mount,
based on your visual judgment. The two holes
for accessing the centering screws (Fig. 4-10/
1
)
are located under the change position.
You can individually position the diaphragm of the iHMC module in the mount, according to the desired
orientation of the relief impression (Fig. 4-9/
2
) in the field of view (Fig. 4-9/
3
). For that, see Examples
A
and
B
in Fig. 4-9.
The concrete approach selected may, of course, deviate from the examples provided. (The position shown
can be altered by as much as 45 degrees.) Whenever iHMC is meant to be employed for several
magnifications, it makes sense to proceed in a similar way to position the diaphragms of the iHMC
modules, as the relief impression will then have the same orientation in all magnifications.
•
Turn the condenser back to its initial orientation until it snaps into position and rotate the clamping
ring to the right for locking.
Fig. 4-10 Centering Screws