INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
Axio Observer
Fitting the microscope stages
ZEISS
12/2016
431004-7244-001
47
4.9.3
Fitting the scanning stage 130x100
STEP
Scanning stage 130x100 STEP rests directly on
three contact points which have drilled screw holes
for fastening.
•
To improve access during stage assembly, the
carrier for transmitted-light illumination
(Fig. 26/
4
) (if fitted) can be tilted backwards.
•
Place the scanning stage 130x100 STEP
(Fig. 26/
1
) on the three contact points
5
) of the stand and fix it in position
using three Allen screws (two at the front
(Fig. 26/
3
), one at the rear (Fig. 26/
2
)).
All stands and stages are supplied with
appropriate spacer disks.
If using flat mounting frames which are
flush to the stage (e.g. 451344) and
160x110 stage pinhole apertures (e.g.
451345-9902) do not use spacer disks
(in case of mechanical and scanning
stages)!
Fig. 26
Fitting the scanning stage 130x100
STEP