When the focusing drive is used in a regular way, a
pop-up window (Fig. 4-28) will appear as soon as
the upper or lower limit stop is reached.
−
Reflected-light illumination
button
1
)
The
Close
and
Open
buttons serve to open
or close the shutter for reflected light (RL).
−
Transmitted-light illumination
button
2
)
The
Close
and
Open
buttons serve to open
or close the shutter for transmitted light (TL).
If there is no shutter in the transmitted-light
path, the voltage of the configured lamp will
be adjusted instead.
−
Make It Visible!
3
)
This button serves to switch the microscope
to a basic state:
−
Transmitted-light lamp adjusted to
medium intensity (2.3 V),
−
Luminous-field diaphragm opened,
−
Aperture diaphragm opened
−
TL shutter open,
−
RL shutter closed,
−
All filter wheels in transmitted light
switched to blank aperture (100%
directed to eyepieces),
−
Condenser switched to brightfield,
−
Reflector turret switched to the nearest
HAL position (Halogen = transmitted
light),
−
Light path switched to 100% to
eyepieces.
Fig. 4-28
Motorized focusing drive has
reached the upper stage limit
switch
Fig. 4-29
Controls