Axio Imager 2
FIRST-TIME SET-UP
ZEISS
Setting the dovetail guide of the stage carrier …
01/2016
430000-7544-001
49
3.19
Setting the dovetail guide of the
stage carrier
When stage carrier 430701-9901-000 is used, the
specimen space needs to be moved for specimens
of less than 13 mm, so that the specimen can be
focused.
To do so, proceed as follows:
•
Remove the stage carrier (see Section 3.17).
•
Loosen
the four clamping screws (Fig. 31/
1
to
4
) of the dovetail guide.
•
Move the guide plate (Fig. 31/
5
) as far
up
as it
will go in the slots.
•
Then, tighten the four screws.
Always
proceed
in the following order:
−
upper left screw (Fig. 31/
1
),
−
lower left screw (Fig. 31/
2
),
−
upper right screw (Fig. 31/
3
),
−
lower right screw (Fig. 31/
4
),
•
Attach the stage carrier (see Section 3.17).
If stage carrier 430701-9901-000 is no longer used, the guide plate must be set back to the
lower
position (normal position).
Fig. 31
Setting the dovetail guide up for
specimen space expansion