Changing the motor
AURA
®
handheld NIR - Rev. 2
Maintenance and Repair Manual | AURA
®
handheld NIR - Rev. 2 | EN_40_070_0030I
6
Remove semicular mount from old motor
Loosen the Torx screw
1
on the semicular mount
2
.
Use a TX 8 Torx screwdriver for this.
Pull the semicular mount
2
off the motor shaft
3
.
Fig. 17
3
2
1
Dismantle the semicular mount
7
Mount the semicular mount onto the new motor
Align the screw
3
of the semicular mount
1
and the flattened side of the
motor shaft
2
so that both point upward during assembly.
Push the semicular mount
1
onto the motor shaft as far as it will go
2
.
Tighten the Torx screw
3
again. Use a TX 8 Torx screwdriver for this.
Fig. 18
2
1
3
Mount the semicular mount