ZEISS
ESSENTIAL OPERATION CONTROLS
Power Supply 12V 100W
6
432610-6044-008
09/2017
Englis
h
9
ESSENTIAL OPERATION
CONTROLS
Connections of the Power Supply unit (rear):
1
Connection socket for reflected light
illuminator (RL)
2
Connection socket for transmitted light
illuminator (TL)
3
Connection socket for control cable
(Remote) to the microscope (Axio Scope.A1
or Axio Examiner)
4
Mains connection
•
Attach the illumination systems (e.g. 2 halogen
lamps HAL 100, Fig. 2/
9
and
11
) to the
transmitted light (Fig. 2/
3
) and reflected light
mount (Fig. 2/
2
), respectively, of the microscope
and fix them using the corresponding clamping
screw (Fig. 2/
1
and
10
).
•
Connect the three-pole lamp plugs (Fig. 2/
7
and
8
) of the two lamps to the rear of the Power
Supply (Fig. 2/
6
): transmitted light illuminator to
the lower socket
TL
, reflected light illuminator
to the upper socket
RL
.
•
If the microscope is provided with a remote
control connection socket (e.g. Axio Scope.A1
or Axio Examiner), connect the microscope
stand with the Power Supply (
Remote
connection socket) using the control cable
(Fig. 2/
5
) for brightness regulation.
Fig. 1
Power Supply (rear)
Fig. 2
Connecting the Power Supply unit