42
AURA handheld NIR – 2289-971 and 2298-955 | User Manual
Appendix
AURA handheld NIR
TECHNICAL DATA SHEET
RRC-SMB-UBC
DS_RRC-SMB-UBC_I
Subject to technical changes without notice * Released document – Copy is not subject to revision service!
2 / 4
Specification
Input
Voltage range
19.00 – 26.00VDC
Current
3.40A max.
Power
65.00W
Output
Voltage range
0 - 17.40VDC
Current range
0 - 4.80A
Voltage tolerance
(1)
±1% max.
Current tolerance
(1)
±10% max. @1.00A, ±3% max. @4.00A
Charge power
60.00W max.
Protection
Short circuit, over temperature shutdown, input-/output over current
Environmental
Cooling
convection cooled
Temperature
Operating:
Non-operating:
0°C to 40°C
-10°C to 70°C
Pressure & Altitude
Operating:
Non-operating:
1060hPa to 795hPa
1060hPa to 572hPa
-382m to 2000m
-382m to 4570m
Humidity
5% to 95% r.H., non-condensing
General
Efficiency
~95% at 100% load
Indicator
Multi-color LED (green, red, orange)
Battery types
Standard battery form factors RRC20x or smart batteries with a similar footprint
Green procurement
RoHS 2011/65/EU
WEEE 2012/19/EU
Chinese RoHS
LED Indications
One time Red/Orange/Green
Self-test: Charger is ready for use.
Red/Green blinking
Battery recognition and initialization.
Orange blinking
The battery is currently being calibrated.
Orange light
The inserted battery is of the correct type and is currently being charged.
Green light
The battery is charged and can be removed for use.
Red blinking
The battery is too hot or too cold to be charged without damage.
If the battery is too cold it will be charged as soon as it has warmed up
sufficiently. If the battery is too hot it should be removed to cool down.
Red light
The battery is damaged or it is a conventional battery which cannot be
recharged.
Charger Mechanical Details
Housing dimensions (LxWxH)
151 x 90 x 43mm
Weight
198g (excluding power supply)
Safety & EMC In combination with included external AC/DC power supply
Regulatory approvals
Europe
CE
International
CB
Electromagnetic Emissions
Europe
EN55011, EN55032, level B
USA
FCC15 class B
Electromagnetic Immunity
ESD immunity
EN/IEC61000-4-2
Electromagnetic field immunity
EN/IEC61000-4-3
EFT / Burst
EN/IEC61000-4-4
Surge
EN/IEC61000-4-5
Conducted Immunity
EN/IEC61000-4-6
Magnetic Fields
EN/IEC61000-4-8
Voltage dips, short instrumentations & volt-
age variations
EN/IEC61000-4-11
Immunity characteristics
EN55024
Содержание 2289-971