ZyPer4K
Electromagnetic Immunity
Test
IEC
Standard
EN60601 Compliance Level
Result Deployment Note
Electrostatic
Discharge (ESD)
61000-4-2
± 6kV contact
± 8kV air
PASS
Expect hard floor of wood, tile,
ceramic, metallic material. If
floor has synthetic covering,
expected relative of humidity of
30% shall be maintained to avoid
static charge buildup.
Radiated RF
61000-4-3
3 V/m
80 MHz to 2.5 GHz
PASS
Minimum separation guidelines
should be observed if placing
equipment close to intentional
RF radiator. Contact ZeeVee for
more details.
Electrical fast
transient/burst
transient
61000-4-4
± 2kV for power supply lines
± 1kV for input/ output lines
PASS
Expected power quality typical
for commercial facility.
Surge
61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
PASS
Expected power quality typical
for commercial facility.
Conducted RF
61000-4-6
3 Vrms
150 kHz to 80 MHz
PASS
Minimum separation guidelines
should be observed if placing
equipment close to intentional
RF radiator. Contact ZeeVee for
more details.
Power frequency
(50/60 Hz)
magnetic field
61000-4-8
3 A/m
PASS
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
61000-4-11 <5%U
T
1
(>95%dipin UT) for
0.5 cycle
40% UT (60% dip in UT) for
5 cycles
70% UT (30% dip in UT) for
25 cycles
< 5% UT (>95% dip in UT)
for 5s
PASS
Typical commercial facility
power expected. Good practice
dictates that a UPS system shall
be employed if continued
operation through any possible
AC delivery disturbance is
desired.