![zebris FDM-T-Sports Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/zebris/fdm-t-sports/fdm-t-sports_operating-instructions-and-specifications_3256047033.webp)
Before connecting or using the measuring system, carry out a visual check of
the power supply unit, power supply cable and socket, as well as the earthing
contacts. Damaged power supply units, cables or plug and socket devices are
to be replaced immediately by an authorized person.
Before connecting the measuring system to the power supply, compare
the nameplate specifi cations on the power supply unit and on the tread-
mill regarding the mains voltage and mains frequency, with the local
characteristics. Connect only if they are compatible.
For operating the FDM-T sensor, exclusively use the power adapter
approved by zebris, which is suitable for the power supply of all
FDM-T systems.
Seite 33/69
Seite 33/41
FDM-T-Sports
5.2. Connection of the measuring system to mains supply
5.2.1. Power supply of the FDM-T sensors
For connecting the FDM-T sensor to the power supply, connect the power supply unit to the mains
socket and the power socket on the interface box.
REF
33102024 / MASCOT Typ 2420
Input
Output Cable
Length
100 – 240 V AC
16 V DC
mains lead
1,7 m
50 - 60Hz
60 W
DC-Leitung
5 m
Pin arrangement/polarity
”
-
” Contact: Outer
contact of hollow plug
”
+
” Contact: Inner
contact of hollow plug