Seite 12/69
Seite 12/41
FDM-T-Sports
3 Product description
3.1. System components
In its basic configuration the measuring system consists of the following components:
■
Treadmill with integrated sensor equipment for measuring the pressure distribution
■
Safety clip for the emergency shutdown
■
Mains cable for connecting the treadmill
■
External power supply unit for the FDM-T pressure plate
■
USB cable (Type A-B)
■
zebris application software zebris FDM
■
Windows-compatible computer or notebook
■
Silicone oil for lubricating the belt
■
Cable guard with screws
■
User manual for FDM-T system, treadmill, zebris FDM software
3.2. Specifications FDM-T sensor
The sensors of the different FDM-T systems only vary in size of the measuring area,
the number of single sensors included in the sensor module and the supported
sampling frequency. All other technical specifications are identical.
Interfaces
USB
synchronization
input/output
videosync.
infrared
synchronization
(optional)
Connectors
interface box on the treadmill housing frame
or connection on the back of the treadmill
Measuring principle
capacitive
pressure
measurement
Operating voltage
16-18
V
DC
Power consumption
maximum 60 W (depending on the type)
Power supply via external power supply unit
100 – 240 VAC/50/60 Hz
Measuring range
1 – 120 N/cm²
Accuracy of the calibrated measuring range
(1 – 80 N/cm
²
) ±5% of maximum range
Hysterese
≤
3%
Pressure threshold
1
N/cm
²