Entering the total test volume (part, tooling and tester) enables a calculated volumetric leak
rate to be shown in the result box in the run mode. This calculation should only be used as a reference.
For a true decay based on a volumetric leak rate, Zaxis manufactures transfer standards to your leak rate
specifications. This ‘leak standard’ is NIST traceable an
d can be used for a daily verification of you test
process.
Pressure/Vacuum Decay Limits
:
A not-to-exceed amount of decay before the timer expires for a pass
condition. If the Decay limit is reached before the timer, the test will
short cycle and result in a fail condition. To see the decay in the full
timer length, check the Evaluate at End of Test box.
An increase limit can be set to verify that no external or thermal
dynamic forces cause the pressure to increase over the test time.
This limit must be enabled by checking the box. If the increase limit
is used, the Evaluate at End of Test box must also be checked.
Occlusion Test Limits
:
A set amount of decay must be reached by the end of the time for a
pass condition. Any decay greater than the limit before the timer
expires will short cycle the test with a pass condition.
To run the test step to the end of the timer regardless of the value
check the Evaluate at End of Test box.
Flow Test Limits
:
A minimum and maximum flow boundary is set for a pass condition.
Pass/Fail condition is evaluated at the end of the Flow step.
Burst/Crack Test Limits
:
A minimum and maximum pressure boundary is set for a pass
condition. The burst trigger is the amount of pressure drop expected
when the burst even occurs.
The test will short cycle when the trigger occurs. The Pass/Fail status
will also be evaluated at the trigger point.