10. Sampler set-up - proportional-to-flow sampling
10.5 Adjust the pump volume adjustment knob to
the value used in the calculation in step 10.4.
10.6 Turn the counter dials to the appropriate number
of pulses you want to count before the sample pump
strokes.
Example: 20 pulses; turn dials to 20.
Press the test button once to load the value into the
memory.
Sample pump displacement per stroke
.1cc
.2cc
.4cc
Number of turns open on the pump volume knob
3
6
12
NOTE: If the calculated counter setting is less than 1 or greater
than 99, the pulses per volume metered will need to be adjusted.
This can be programmed in most flow meters to the desired rate.
If the calculated counter setting is less than 1, increase the pulses
per volume metered. If the calculated counter setting is greater
than 99, decrease the pulses per volume metered.