background image

Maintenance and Storage

15

16

Figure 11

Figure 12

Figure 10

Clean the surface of the fingertip oximeter before it is used in diagnosis for patients.
The most commonly used hospital cleaning and non-corrosive liquid detergent can be used to clear 
the  oximeter.  Pay  attention  to  diluting  cleaning  detergent  cleaning  before  use,  by  following  the 
manufacturer's instructions.
Avoid the use of ethanol-based, amino-or acetone-based cleaning agent.
Oximeter shell should be maintained from dust pollution, using a soft cloth or lint-free cleaning agent 

with the sponge to wipe infiltration. Make sure no liquid inside the equipment.
Avoid  the  use  of  metals  such  as  steel  wire  brush  or  polishing  agent  abrasive  material  which  will 
damage the oximeter panel .
Under normal conditions there in no need for special protection and maintenance when using, please 
pay attention to the following points:
Using oximeter in required environment.
Avoid direct sunlight exposure.
Avoid extreme infrared radiation or ultraviolet radiation.
Avoid organic solvent vapors, dust, and corrosive gas exposure.
Warning: Do not put oximeter in disinfection or in liquid.
Please remove the battery if the oximeter will not be used for long time.
Transportation and storage conditions:
Temperature range: -20℃~+55℃                Relative humidity: ≤93%, no condensation 
Atmosphere pressure: 500hPa~1060hPa

Possible cases and solutions

Warning: 
Oximeter cover can only be opened by a professional maintenance staff. No internal parts require 
opening by end users.
If you are not sure about the measurement precision, please use other methods to check patient's 
pulse, to determine whether oximeter works.
Note:  Do  not  splash,  dump  any  liquid  into  the  oximeter  and  attachments,  switch  and  connections 
which may damage the oximeter.

19

20

1. Normal

2. Replace the batteries

1. The product is automatically powered 

off when no signal is detected longer 

than 8 seconds
2. Finger quantity of the batteries is start 

being inadequate

Indication 

lamp are 

suddenly 

off 

Problems  

            Possible reason                                                                                    Solution

3. Nail polish or paste Manicure

1. Do not put finger correctly

2. Try some more times, if you can make sure about no problem exiting 

in the product, please go to a hospital timely for exact diagnosis
3. When measuring the nail polish remover or discharge Manicure

2. Patient Oxyhemoglobin value is too 

low to be measured

1. Try again

SPO

2

 or PR 

can not be 

shown 

normally

1. Retry by plugging the finger

2. Try not to move

2. Finger is trembling or patients body is in 

movement status

1. Please replace batteries

2. Please reinstall the batteries

SPO

or PR 

is shown 

unstably 

1. Finger might not be plugged deep 

enough

1. Power of batteries might be inadequate 

or not be there at all
2. Batteries might be installed incorrectly

3. The Oxyhemoglobin might be damage 3. Please contact with local customer service center

The 

Oxyhemog

lobin can 

not be 

power on

Содержание YX3 Series

Страница 1: ...ity Pulse Scanning and Recording Technology so that two beams of different wavelength of light 660nm glow and 940nm near infrared light can be focused onto human nail tip through perspective clamp finger type 9 Warning The water protection grade of device is IPX 1 10 Warning Please take out the batteries and do not burn or bury if the product is scrapped or discarded 11 Warning Device conforms to ...

Страница 2: ...tronic apparatus 5 High definition television In this interference the measurement values may deviate or it does not work When interfered the product may produce abnormal performances unstable reading values outages or other functions of error If such a case the use of the site should be checked to identify interference and the elimination of the following measures 1 The vicinity of the equipment ...

Страница 3: ... the whole body be in silent status 6 Read corresponding data form display screen Product scope of application Contraindications None Intended Use The Fingertip Pulse Oximete is a kind of non invasive detection adults and children spo2 and heart rate detection of medical instrument which is expected for home and hospital inspection Applicable People Normal adults and children Vascular disease crow...

Страница 4: ...m 34mm LWH Weight 54g approximately including two AAA batteries 9 Working Environments Ambient Temperature 5 40 Relative humidity 80 Atmospheric pressure 860hPa 1060hPa Transportation And Storage Environments 40 55 OLED Use 20 55 Relative humidity 93 Atmospheric pressure 500hPa 1060hPa 10 Operation mode Continue operation 11 Patient population The personal application are adult patient weight 30kg...

Страница 5: ...a 1060hPa Possible cases and solutions Warning Oximeter cover can only be opened by a professional maintenance staff No internal parts require opening by end users If you are not sure about the measurement precision please use other methods to check patient s pulse to determine whether oximeter works Note Do not splash dump any liquid into the oximeter and attachments switch and connections which ...

Страница 6: ...y stopworking interfered by strong radio frequency or because of close distance Medical places may have thefollowing EMI pollution sources 1 Electronic surgical instruments 2 Mobile phones 3 Automotive two way wireless communications equipment 4 Electronic apparatus 5 High definition television In the designed oxygen using environment EMI may make the pulse not obvious In this interference themeas...

Страница 7: ...ESD IEC61000 4 2 Guidance and MANUFACTURER S declaration ELECTROMAGNETIC IMMUNITY for ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Guidance and manufacture s declaration electromagnetic immunity IMMUNITY test IEC60601 test level Compliance level Electromagnetic environment guidance 3V m 80MHz to 2 5GHz 3V m Radiated RF IEC61000 4 3 The Finger Pulse Oximeter YX300 YX301 YX302 is intende...

Страница 8: ...YunYang Industrial Park DanYang Jiangsu Province P R China 212300 http www yuwell com Warranty Card 1 Fingertip Pulse Oximeter 1pc 2 Lanyard 1pc 3 AAA batteries 2pcs 4 User s Manual Warranty card 1pc Of waste waste handling should be consistent with the corresponding national laws and regulations Waste residue from processing NOTE1 At 80MHz and 800MHz the higher frequency range applies NOT2 These ...