Product Properties

Figure 5: Testing sketch diagram

1.Operation of the product is simple and convenient.
2.The product is small in size, light in weight (total weight is about 54g including batteries) and portable.
3.The power of two AAA batteries can last for 45 hours.
4.Low voltage warning will be indicated in visual window when battery voltage is too low than normal 
operation of the oximeter might be influenced.
5.The product will automatically shut down when there is no finger entering for eight seconds.

07

08

11

12

Display

Product Operation Scope

The fingertip Oximeter can be used to measure human Haemoglobin Saturation and heart rate through 
finger,  the  product  is  suitable  forfamily,  hospital  (including  clinical  use  in  internist/surgery,  Anaesthesia, 
paediatrics, intensive sports (It can be used before or aftersport. Operation in sport is not recommended) 
and etc.
The product is not suitable to monitor patient continuously.

Product Operation Scope

1.Installing two AAA batteries into battery cassette before closing the cover.
2.Nip the clamp as diagram.
3.Plug one finger into rubber hole of the Oximeter (it is best to plug the finger 
thoroughly) before releasing the clamp.
4.Press the switch button one time on the front panel.
5.Do  not  tremble  while  the  oximeter  is  working.  It's  better  that  the  whole 
body be in silent status.
6.Read corresponding data form display screen.

Product scope of application

Contraindications: None.
Intended Use: The Fingertip Pulse Oximete is a kind of non-invasive detection adults and children spo

and heart-rate detection of medical instrument which is expected for home and hospital inspection.
Applicable  People:Normal  adults  and  children,Vascular  disease  crowd,The  respiratory  system 
disease crowd,Middle-aged and old old man over the age of 60,athlete.

Figure 2  Button mode

Symbol of Oxygen saturation

Battery level 

Pulse strength

Pulse rate

Pulse rate

Pulse volume

wave

Pulse volume 

wave

Oxygen

saturation

Oxygen

saturation

9 9

75

SpO

2

%

P R bpm

Symbol of Oxygen saturation

Pulsestrength

Figure 3  Gravity sensing mode 

Battery level 

Symbol of PR

Symbol of PR

Symbol of Gravity 

sensing mode 

9 9

75

SpO

2

%

P R bpm

Technical Parameters

1.Display Type: OLED 
SPO

2

 Measurement rang: 70-100%, Accuracy: ±2%, Pulse Rate Measure rang: 25-250BPM/min,

Accuracy:±1BPM or±1% (larger) 
2.Power: Two AAA 1.5V alkaline Batteries, Voltage range: 2.6-3.6V  
3.Working Current: Less than 30mA.
4.Measurement accuracy:

Figure 4 Wave wrong mode

OLED display wrong mode when finger out or product failure.

Wrong Mode Display

SpO

2

%

P R bpm

acquired  through  which  will  be  shown  on  screen  through  treatment  in  electronic  circuits  and 
microprocessor. 
The pulse oximeter is intended for use in hospital, clinical institution, healthcare community
1) data averaging and other signal processing
2) the data update period

Содержание YX3 Series

Страница 1: ...ity Pulse Scanning and Recording Technology so that two beams of different wavelength of light 660nm glow and 940nm near infrared light can be focused onto human nail tip through perspective clamp finger type 9 Warning The water protection grade of device is IPX 1 10 Warning Please take out the batteries and do not burn or bury if the product is scrapped or discarded 11 Warning Device conforms to ...

Страница 2: ...tronic apparatus 5 High definition television In this interference the measurement values may deviate or it does not work When interfered the product may produce abnormal performances unstable reading values outages or other functions of error If such a case the use of the site should be checked to identify interference and the elimination of the following measures 1 The vicinity of the equipment ...

Страница 3: ... the whole body be in silent status 6 Read corresponding data form display screen Product scope of application Contraindications None Intended Use The Fingertip Pulse Oximete is a kind of non invasive detection adults and children spo2 and heart rate detection of medical instrument which is expected for home and hospital inspection Applicable People Normal adults and children Vascular disease crow...

Страница 4: ...m 34mm LWH Weight 54g approximately including two AAA batteries 9 Working Environments Ambient Temperature 5 40 Relative humidity 80 Atmospheric pressure 860hPa 1060hPa Transportation And Storage Environments 40 55 OLED Use 20 55 Relative humidity 93 Atmospheric pressure 500hPa 1060hPa 10 Operation mode Continue operation 11 Patient population The personal application are adult patient weight 30kg...

Страница 5: ...a 1060hPa Possible cases and solutions Warning Oximeter cover can only be opened by a professional maintenance staff No internal parts require opening by end users If you are not sure about the measurement precision please use other methods to check patient s pulse to determine whether oximeter works Note Do not splash dump any liquid into the oximeter and attachments switch and connections which ...

Страница 6: ...y stopworking interfered by strong radio frequency or because of close distance Medical places may have thefollowing EMI pollution sources 1 Electronic surgical instruments 2 Mobile phones 3 Automotive two way wireless communications equipment 4 Electronic apparatus 5 High definition television In the designed oxygen using environment EMI may make the pulse not obvious In this interference themeas...

Страница 7: ...ESD IEC61000 4 2 Guidance and MANUFACTURER S declaration ELECTROMAGNETIC IMMUNITY for ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Guidance and manufacture s declaration electromagnetic immunity IMMUNITY test IEC60601 test level Compliance level Electromagnetic environment guidance 3V m 80MHz to 2 5GHz 3V m Radiated RF IEC61000 4 3 The Finger Pulse Oximeter YX300 YX301 YX302 is intende...

Страница 8: ...YunYang Industrial Park DanYang Jiangsu Province P R China 212300 http www yuwell com Warranty Card 1 Fingertip Pulse Oximeter 1pc 2 Lanyard 1pc 3 AAA batteries 2pcs 4 User s Manual Warranty card 1pc Of waste waste handling should be consistent with the corresponding national laws and regulations Waste residue from processing NOTE1 At 80MHz and 800MHz the higher frequency range applies NOT2 These ...

Отзывы: