3.6 Operation of B-scan Interface

3.6.1 B-scan Display

3.15 B scan interface diagram

1-B-scan image display 2-White pointer

3-Red triangle (display the min.

thickness value)

4-The thickness value of the pointer position

5-The

minimum thickness value in B-scan 6-The maximum thickness value in B-scan
7-Erasing the current B-scan images and measurements 8-Enter the numerical
measurement interface 9-Sound velocity 10-The minimum thickness value on
the B-Scan image 11.Parameter display area 12.Gain value

3.6.2 Introduction of B-Scan

UM-5 Series thickness gauge has time-base B-scan function. Move the probe
along the workpiece surface, then the cross-sectional profile of the workpiece
display, use for observe the underside contour of the workpiece.
When remove the probe from workpiece, it could be automatically capture a
minimum value of the B-scan image, and indicate the position of the
minimum by a red triangle. The thickness value of any point in the B-scan
image can be viewed by moving the white pointer.

3.7 Dual-echo (Thru-coating) Mode

Having coating or paint layer on the surface of the workpiece, it will bring
considerable error to the measurement results. UM-5D and UM-5DL have the
echo-echo measurement method, which can accurately measure the actual
thickness of the substrate under the workpiece coating without grinding the
coating and other processes that destroy the workpiece surface. This function
is achieved by measuring two successive bottom echoes of the substrate.
Press

into parameter configuration interface, set the measurement mode

to dual-echo and press

again back to thickness measurement interface.

And then user can measure the thickness through coating.

3.7.1 A-scan Interface in Dual-echo Mode

The menu options on the right side have changed of A-scan interface under
echo-echo mode, added E-blanking option, and canceled GATE option. The
blue strips area indicate the length of echo-blanking when measuring, the
waveform above the blue strips is invalid. See following figure 3.16.

Figure 3.16 A-scan interface in dual-echo mode

1. identification of dual-echo measurement mode

2. Blue arrow indicate the

secondary echo 3. E-blanking

4. Blue line: the length of echo-blanking

5. Red line: the length of initial-blanking

6. Red arrow indicate the first echo

Blanking in the dual-echo mode:

19

20

Содержание UM-5 Series

Страница 1: ...15 3 5 Real Cases Analysis of A Scan 17 3 6 Operation of B scan Interface 19 3 6 1 B scan Display 19 3 6 2 Introduction of B Scan 19 3 7 Dual echo Thru coating Mode 20 3 7 1 A scan Interface in Dual...

Страница 2: ...splay is available by showing the echoes which enables the more extensive measure range more precise result and very little measurement error and finally makes the measuring processes more controllabl...

Страница 3: ...LF FULL Display Mode Thickness Value Min Max capture DIFF RR A Scan B Scan V Path Correction Automatic compensate the non linearity of dual element probe Update Rate Selectable 4Hz 8Hz 16Hz Material V...

Страница 4: ...eat capacity data storage function Stores 100 000 thickness values 1000 waveforms only UM 5DL 15 Measure the net thickness of the workpiece through the coating layer only UM 5D UM 5DL 2 Keyboard Funct...

Страница 5: ...If the material velocity is known for example the velocity of steel is 5900m s The sound velocity can be setting manually 3 1 1 Probe zero calibration Note The calibration result is 4 00mm only under...

Страница 6: ...on adjusting options like FILE NUMBER MEASURING MODE VIEW MODE PROBE SETUP MINIMUM ALARM MAXIMUM ALARM NORMAL THICKNESS THE MINIMUM OF B SCAN THE MAXIMUM OF B SCAN RECTIFICATION RECTIFICATION WAVEFORM...

Страница 7: ...TIFICATION MODE RF full wave half half RF RF describes the complete echo waveform full wave indicates the half echo and the overturned half echo half means putting off the half echo and turn the half...

Страница 8: ...erence between measured and nominal thickness and reduction difference as a percentage of nominal thickness as well as the current measured and nominal thickness values Before using this mode presetti...

Страница 9: ...mall increments while left and right for large increments GAIN Adjust the amplification of echo signal by instrument to allow manual increase or decrease in ldB This function is very useful for the te...

Страница 10: ...echo brake through the gate to achieve the correct measurement 3 12 REAL CASE 1 2 There are some defects in the testing object and the gate locks the defect echoes As showed in the following figure th...

Страница 11: ...ng or paint layer on the surface of the workpiece it will bring considerable error to the measurement results UM 5D and UM 5DL have the echo echo measurement method which can accurately measure the ac...

Страница 12: ...kness value 7 delete the selected data 8 The current cell selected Whether it is in the interface of thickness value A scan or B scan the current measured thickness values will be stored for short pre...

Страница 13: ...t 5 2 Measuring Methods 1 SINGLE POINT MEASUREMENT Using the probe to measure a random point on the surface of the object the reading displayed is the thickness value 2 DOUBLE POINT MEASUREMENT Measur...

Страница 14: ...time the gauge will not be turned on after running out of power 6 2Considerations 1 Please be cautious of the zero block s getting rust as couplant will be spread on the surface of it when calibrating...

Страница 15: ...5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0 110 1700 Polyethylene...

Отзывы: