2.2 D Type Impact Device Diagram

图 2.2

Figure 2.2

2.3 Types of Impact Device

Figure 2.3

2.4 Impact Devices Specification:

Special Impact

Device

D/DC/DL

D+15

C

G

E

Impact Energy

11mJ

11mJ

11mJ

11mJ

11mJ

Mass of Impact Body

5.5g/7.2g

7.8g

3.0g

20.0g

5.5g

Test Tip Hardness

1600HV

1600HV

1600HV

1600HV

5000HV

Test Tip Diameter

3mm

3mm

3mm

5mm

3mm

Test Tip Material

Tungsten

Carbide

Tungsten

Carbide

Tungsten

Carbide

Tungsten

Carbide

Tungsten

Carbide

Impact.Device Diameter

20mm

20mm

20mm

30mm

20mm

Impact Device Length

86/147/75

162mm

141mm

254mm

155mm

Impact Device Weight

50g

80g

75g

250g

80g

Max.Hardness

of

Sample

940HV

940HV

1000HV

650HV

1200HV

3

4

Содержание LM-100

Страница 1: ...Leeb Hardness Tester LM100 Operating Manual YUSHI INSTRUMENTS...

Страница 2: ...ss Requirements 9 4 1 3 Surface Hardened Layer Thickness 9 4 1 4 Curved Surface Requirement 9 4 2 Support of Test Piece 10 5 Operation 10 5 1 Display 10 5 2 Keyboard 11 5 3 Turn on the Instrument 11 5...

Страница 3: ...uce directly proportional voltage with the impact velocity The Leeb hardness values are calculated by the following formula HL 1000 Vb Va In Which HL Leeb hardness values Vb the voltage produced durin...

Страница 4: ...7 2g 7 8g 3 0g 20 0g 5 5g Test Tip Hardness 1600HV 1600HV 1600HV 1600HV 5000HV Test Tip Diameter 3mm 3mm 3mm 5mm 3mm Test Tip Material Tungsten Carbide Tungsten Carbide Tungsten Carbide Tungsten Carbi...

Страница 5: ...7 m 10 m Impact Device Applicable Scope DC Used for very confined spaces such as holes cylinders internal measurement DL Used for extremely confined spaces D 15 Used for measuring in grooves recessed...

Страница 6: ...D type impact device Standard Optional C G DC DL D 15 Accuracy 6HLD HLD 800 Repeatability 6HLD HLD 800 Measuring Direction Vertically downward inclined downward Level inclined upward vertically upwar...

Страница 7: ...ce to be tested is smaller than 30mm D DC D 15 C E and DL type of impact device and smaller than 50mm G type of impact device the small support ring or the special support rings should be selected 4 2...

Страница 8: ...e button to select the probes 5 4 4 Impact Direction Press DIR button to select impact direction Figure 5 4 5 4 5 Impact Times Press TIME button to set the impact times from 1 7 which will calculate t...

Страница 9: ...7 Data Save and Read Setting 5 7 1 Storage Testing Result Press MEM storage button to set the saving function the S symbol will be shown on the display The tester will save the testing results automat...

Страница 10: ...is will cause the measuring error so we design a calibration procedure of using the standard testing block to calibrate Press up and down button at the same time after the power on then the tester wil...