background image

Pull back the cocking barrel until 
you hear a click. Now the lancing 
device is loaded and ready to 
draw blood.

Before taking a blood sample, 
wash your hands or use an 
alcohol swab to clean the area. 
Washing your hands in hot water 
increases blood circulation.  You 
can also massage from wrist to 
finger to promote better blood 
circulation.

Holding the lancing device 
against the side of the finger to 
be lanced, press the release 
button and then put down the 
lancing device. Massage 
forward slowly from the base 
of your finger to the tip to 
increase the sample size. 

 

Wipe away the first drop of 
blood and use the second drop 
for the test strip.

Tip:

 

To reduce pain, lance on 

the sides of the fingertips, 
where there are less nerve 
endings. Rotate finger 
locations as much as 
possible to accelerate 
wound healing and decrease 
callouses.

      

Forearm or Palm Testing

Collecting blood samples from your palm or forearm may be less painful 
than your fingertips because there are less nerve endings. The procedure 
for collecting samples from your palm or forearm is slightly different.  You 
will need the clear cap for this type of sampling. You cannot adjust the 
puncture depth with the clear cap.

IMPORTANT: There are important differences among forearm, palm 
and fingertip samples that you should know. Important information 
about forearm, palm glucose testing:

Please refer to "Fingertip Testing" sections 1-3 and 5 to learn how to install 
lancets into lancets holder.

When blood levels are changing rapidly such as after a meal, 
insulin dose or exercise, blood from the fingertips may show 
these changes more rapidly than blood from other areas.
Fingertips should be used if testing is within 2 hours of a meal, 
insulin dose or exercise and any time you feel glucose levels are 
changing rapidly.
You should test with the fingertips anytime there is a concern for 
hypoglycemia or you suffer from hypoglycemia unawareness.

Screw the clear cap onto the 
lancing device.

Choose a sampling site on 
your forearm or palm that is 
dry and clean. Do not choose 
an area near bone, visible 
veins, or hair. To bring fresh 
blood to the surface for 
sampling, massage the site 
vigorously until it feels warm.

Place the lancing device 
against the puncture site. Press 
and hold the clear cap against 
the puncture site for a few 
seconds. Press the release 
button of the lancing device, 
but do not immediately lift the 
lancing device from the 
puncture site. Continue to hold 
the lancing device against the 
puncture site until you can 
confirm a sufficient blood 
sample has formed.

Содержание GoChek

Страница 1: ... Haishu Road Yuhang District Hangzhou Zhejiang 311121 China Phone 86 571 88566372 Website http www microtechmd com EC REP Lotus Global Co Ltd 1 Four Seasons Terrace West Drayton Middlesex London UB7 9GG United Kingdom BLOOD GLUCOSE MONITORING SYSTEM Instruction Manual ...

Страница 2: ... meter or get it wet Only use the meter as directed by this instruction manual Only use GoChek blood glucose testing strips hereafter referred to as test strips Do not use the meter if is not working properly or has been damaged Do not place anything on top of the meter Unless specifically instructed in this user manual do not let anything get inside of the meter openings This meter can only be us...

Страница 3: ...ting Humidity 20 90 non condensing Hematocrit Range 20 70 Note Blood containing high levels of vitamin C or other reducing agents can lead to inaccurate test results Triglycerides above 3 000 mg dL and cholesterol above 500 mg dL will lead to inaccurate test results Specification Product Components Component Descriptions Blood Glucose Meter Display Test Strips Code Chip Control Solution Meter Setu...

Страница 4: ...Button When the meter is off holding the M button will enter memory mode This button also is used to select confirm Button When the meter is off holding the button will enter the test reminder alarm mode Battery Door Remove the battery door to install one CR2032 coin cell battery Label Contains product information Test Strip Ejector Slide the ejector forward to discard a used test strip Data Port ...

Страница 5: ...le is applied at the edge of the test strip never on the top surface Applying blood to the top surface of the strip will lead to inaccurate results Apply blood samples to the edge of the strip until the test window is full The blood glucose meter will begin to count down when there is adequate blood If the test window is not full you may add additional blood within three seconds If not enough bloo...

Страница 6: ...s for more detailed information Code Chip Insert the code chip into the meter before each test It is used to calibrate the meter for use with the associated test strips Control Solution Control solution is a glucose solution of known concentration that is used to confirm that your meter and test strips are working properly It is important to run a quality control test regularly to make sure that y...

Страница 7: ...sound Press the M button to save this setting When the sound is on the meter will provide a beep when You will hear a one beep warning sound when the meter is powered on if the blood sample is adequately applied and when the test results are displayed You will hear 2 short beep warning sounds if an error occurs Please refer to the error code table included in this instruction manual After you have...

Страница 8: ... upper right section of the display if your test result is higher than 16 7 mmol L 300 mg dL After choosing the ketone test warning setting the system will enter the meal marker setting Meal Marker Settings The meter has a meal marker function that allows you to mark your test result as before or after a meal To turn on or off meal marker function press the or buttons and then the M button to save...

Страница 9: ...operly When you suspect that your test results are inaccurate or if they are inconsistent with the way you feel When you suspect the meter is damaged Insert the code chip into glucose meter Insert the test strip all of the way into the test strip port and the meter will turn on automatically If the audio is on you will hear a beep sound and all of the icons on the display will turn on at the same ...

Страница 10: ...olution 1 and Control Solution 2 Confirm you are matching the correct range Control Solution 1 results should be matched to the CTRL1 range printed on the test strip vial or foil pouch Check the expiration date of the test strip and control solution Make sure that the packages have not been opened for more than 6 months Discard any expired test strips and control solution Confirm that you are test...

Страница 11: ...ower off After the meter detects that there is an adequate amount of blood the meter will count down 5 seconds and display the glucose test results The test result will be stored in the history automatically Slide the test strip ejector to eject the test strip and the meter shuts off automatically Clear Cap Use for sampling on arm or hands Lancet Release Button Safety Cap Lancing Device Cap Depth ...

Страница 12: ...h the clear cap IMPORTANT There are important differences among forearm palm and fingertip samples that you should know Important information about forearm palm glucose testing Please refer to Fingertip Testing sections 1 3 and 5 to learn how to install lancets into lancets holder When blood levels are changing rapidly such as after a meal insulin dose or exercise blood from the fingertips may sho...

Страница 13: ...ot immerse the lancing device in water Control excess bleeding and disinfect your wound after use Reminder Testing Blood Glucose When you insert the code chip and test strip the meter will automatically turn on except in data transfer mode Insert the code chip into glucose meter Insert the test strip all of the way into the test strip port and the meter will turn on automatically If the audio is o...

Страница 14: ...g is on you can press the button to tag the reading before or after meal as shown in the figure Press the M button to confirm DO NOT The display will count down 5 seconds during the measurement process If the audio is on the test will end with a beep If you applied a blood sample but the meter does not begin a countdown you may reapply a second drop of blood within 3 seconds If an error code is di...

Страница 15: ...LO messages indicate that the test measurement was outside of this range If HI is displayed it indicates the test result is above 33 3 mmol L 600 mg dL You should test again to ensure that no mistake was made in the procedure If you are sure your glucose meter is functioning properly no mistakes were made and the test results still show HI then you may be in a state of severe hyperglycemia please ...

Страница 16: ...self monitoring The test results can t be referred to as confirmed cases In order to ensure the accuracy of the results the test results can be further confirmed by other methods such as biochemical methods As with all diagnostic reagents test results must be linked with a professional doctor s diagnosis from the other clinical symptoms Carefully process waste caused by blood glucose tests accordi...

Страница 17: ...ocal government Comparing Meter and Laboratory Results Make sure your meter is working properly Comparisons will be more accurate if you do not eat for at least four hours preferably eight hours before testing Bring your blood glucose meter test strips and control solution to the lab Ensure that the time between tests with your meter and the laboratory is within 15 minutes Wash and dry your hands ...

Страница 18: ...ren and pets Do not disassemble the meter Disassembly will void the warranty Please follow the local regulations to dispose of the meter and battery properly Make sure the meter is off Open the battery door on the back of the meter Remove the old battery and replace with a new CR 2032 3 0V coin cell battery making sure that the positive side is facing up Replace the battery door After replacing th...

Страница 19: ...est strip is ok remove the battery for 30 seconds and retest with a new strip If the problem persists contact customer service Retest with a new strip Make sure there is enough blood to fill the test window Repeat the test and ensure test strip remains in place Move to a place within the normal operating temperature range and repeat the test The next 20 results will still be accurate but replace t...

Отзывы: