background image

6

7

It is not necessary to calibrate in both % and mg/L or ppm. 

Calibrating in % will simultaneously calibrate 

mg/L and ppm and vice versa.  

CALIBRATING DO% IN WATER-SATURATED AIR:  

1-POinT CalibraTiOn

Make sure there are no water droplets on the OBOD sensor cap or temperature 

sensor and then place the probe in a BOD bottle that contains a small amount 

of clean water (approximately 40 ml).  The BOD bottle should be clean since 

bacterial growth may consume or produce oxygen which would interfere with 

the calibration.  Make sure the DO and temperature sensors are not immersed in 

water.  The purpose is to create a 100% water-saturated air environment.  Next, 

wait approximately 5 to 10 minutes or until the DO and temperature readings 

are stable.  This will allow for the air in the BOD bottle to become completely 

water-saturated.    The  design  of  the  probe  ensures  venting  to  the  atmosphere 

when placed in a typical BOD bottle which is required to create the 100% water-

saturated air calibration environment.     

Press the Calibration 

 key on the ProODO instrument.  Highlight DO and 

press enter.  Highlight 

DO % 

and press enter to confirm. 

The  instrument  will  use  the  reading  from  the 

internal barometer during calibration and will 

display this value in brackets at the top of the 

display.    Highlight  the  barometer  value  and 

press enter to adjust it if needed.  Note  - the 

barometer should be reading “true” barometric 

pressure.  If the barometer reading is incorrect, 

perform  a  barometer  calibration.    See  the 

Barometer section in the ProODO User Manual 

for  more  information  on  “true”  barometric 

pressure  and  on  how  to  perform  a  barometer 

calibration.  If the barometer reading is acceptable, there is no need to change it 

or perform a barometer calibration.   

 

Wait for the temperature and DO% values under “Actual Readings” to stabilize, 

then highlight 

accept Calibration 

and press enter to calibrate.  If User Field 1 

or 2 are enabled, you will be prompted to select the fields and then press Cal 

to complete the calibration.  The message line at the bottom of the screen will 

display “Calibrating Channel...” followed by “Calibration Successful”.  Press Esc 

 to cancel the calibration. 

i

temperatUre

All ProOBOD probes have built-in temperature sensors.  Temperature calibration 
is not required nor is it available.  To set the units, press the Probe 

 key on 

the ProODO instrument and select 

Display

.  Highlight 

Temperature

 and press 

enter.  Highlight the desired temperature units of 

°F

,

 °C

, or 

K

 and press enter to 

confirm the selection.   

taking measUrements

Important:  The ProOBOD is designed for laboratory use, not 

field use.  It is not waterproof and should never be immersed 

in a sample past the taper on the stem of the probe

To take readings, insert the probe into the BOD sample bottle.  The DO and 

1. 

temperature sensors should be immersed in the sample.  

Turn the probe stirrer on with the toggle switch located on the top of the 

2. 

probe (Figure 1).  The stirrer helps keep solids from settling at the bottom 

of the sample bottle and quickens the sensor’s response.

Allow the temperature readings to stabilize and wait approximately 20-25 

3. 

seconds for the DO readings to stabilize.  

Log One Sample will be highlighted on the ProODO instrument when it is 

4. 

in Run mode.  Press Enter to open a logging submenu.  Highlight Sites or 

Folders and press Enter to select a site or folder to log the sample to. 

Once the Site and/or Folder name is selected, highlight 

5. 

log now

 and press 

Enter.    The  instrument  will  confirm  that  the  data  point  was  successfully 

logged.   

If you would like to log at a specific interval vs. logging one sample at a time 

6. 

or vice versa, press the 

System

 

key, then highlight 

logging

 and press 

Enter.  Select 

Continuous Mode

 and adjust the time interval if necessary.    

On the Run screen, the option to log will change from Log One Sample to 

Start Logging based on the time interval entered in the Logging Menu.  

Note – When utilizing the Manual sampling mode, the  

Continuous Logging Interval must be set to 10 seconds or 

greater.   For additional information on logging data and the 

sampling mode options, refer to the ProODO User manual.

u

i

Содержание WE KNOW ProOBOD

Страница 1: ...USER MANUAL ...

Страница 2: ...ntenance and Storage 8 Updating Instrument and Probe Firmware 8 Sensor Maintenance Dissolved Oxygen 8 Sensor Maintenance Temperature 11 Sensor Storage 12 Troubleshooting 12 Dissolved Oxygen Readings 12 Help 13 Error Status messages 13 Restore Default Calibration Values 14 Probe Specifications 15 Accessories Parts List 16 Declaration of Conformity 17 Recycling 19 Contact Information 20 Ordering and...

Страница 3: ...elected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by Failure to install operate or use the product in accorda...

Страница 4: ...tional outlet adapters are also included The ProOBOD sensor measures dissolved oxygen in water using lifetime luminescence technology and uses a digital signal to send information between the instrument and probe The probe features a stirrer and motor to aid in sample movement to keep solids from settling at the bottom of the sample bottle and to speed up the response of the sensor Key advantages ...

Страница 5: ...nstruction sheet is important because it contains the calibration coefficients for your sensor cap After using this sheet for general probe setup be sure to store it in a safe place in case you need to reload these calibration coefficients in the unlikely event that they are ever deleted from the instrument and probe Note A new cable probe assembly already has a sensor cap installed and the sensor...

Страница 6: ...r cap needs replaced See the Sensor Maintenance Dissolved Oxygen section of this manual or the instruction sheet included with the replacement sensor cap for instructions on updating the Sensor Cap Coefficients When updating the Coefficients the sensor cap serial will be updated automatically based on your entries The Temperature Coefficient should not be modified unless instructed by YSI Technica...

Страница 7: ...nd then press Cal to complete the calibration The message line at the bottom of the screen will display Calibrating Channel followed by Calibration Successful Press Esc to cancel the calibration i Temperature All ProOBOD probes have built in temperature sensors Temperature calibration is not required nor is it available To set the units press the Probe key on the ProODO instrument and select Displ...

Страница 8: ...en replacing a Sensor Cap the Sensor Cap Coefficients must be manually updated in the ProODO instrument The instruction sheet shipped with the replacement OBOD sensor cap includes the calibration coefficients that are specific to your Sensor Cap u IMPORTANT Be sure to save the OBOD Sensor Cap instruction sheet in case you need to reload the calibration coefficients The replacement OBOD Sensor Cap ...

Страница 9: ...Calibration Code has been entered correctly press enter to confirm Highlight K2 through KC in turn and use the 5 numeric entry screen to enter the corresponding new coefficient from the Calibration Code label as described in step 4 Press enter after each entry and then proceed to the next K selection After all the new coefficients have been entered highlight 6 Update Coefficients and press enter A...

Страница 10: ... sensor cap and then recalibrate If you are still getting erroneous dissolved oxygen readings try replacing the sensor cap and then recalibrate If the erroneous readings continue contact YSI Technical Support to help determine the next step Inspecting the Sensor Cap for Damage Caution Avoid touching the sensing end of the sensor cap during the following maintenance procedures If DO measurements se...

Страница 11: ... Values Occasionally the instrument may need to have the factory calibration default values restored To perform a factory reset press Calibrate highlight Restore Default Cal and press enter Select the parameter you wish to restore either DO or Barometer and press enter After selecting barometer or DO you will be asked to confirm the operation Highlight Yes and press enter P robe S pecifications Th...

Отзывы: