background image

Measuring discharge - Procedure 

 

Select a site with reasonable uniform bottom conditions 

that may be safely waded (typically no more than 1-m 

deep). String a graduated tag line across the river 

perpendicular to the riverbanks. 

Start at one edge and enter the starting edge location and 

water depth. Divide the river cross-section into a number 
of stations appropriate for its width. In the above diagram, 

the river has start (Loc

0

) and end (Loc

12

) edges with 11 

stations (marked Loc

1

 to Loc

11

) in between. 

At each station, the operator enters the station location 

and water depth and measures the velocity at one or more 

depths. A variety of measurement methods are supported 

using measurements at different depths, including: 

 

0.6 * Depth 

 

0.2 and 0.8 * Depth 

 

0.2, 0.6 and 0.8 * Depth 

 

Kreps (surface and 0.62 * Depth) 

 

5-point (surface, bottom, and 0.2/0.6/0.8 * Depth) 

 

Multi-point method (any number of points at user 

specified depths) 

During the entire measurement, the probe’s X-axis must 

be maintained perpendicular to the tag line. The probe 

should be held away from underwater obstacles that may 

disturb the flow. Do not turn the FlowTracker into the 

direction of flow, as it will automatically account for flow 

direction when making discharge measurements. 

Sampling 

Volume

Y

X

Probe

Coordinate System

Graduated Tag Line

Primary Flow

Direction

Mounting

Pin

 

Orient the FlowTracker perpendicular to the tagline 

Velocity data is recorded once per second for the entire 

Averaging Time, and then averaged to compute the mean 

velocity. Quality control data is also reviewed and 

displayed; you will be alerted to any unexpected values. If 

the velocity measurement is found to be unsatisfactory, 

you should repeat the measurement. 

Based on the depth, station width, and mean velocity, the 

discharge for a station is calculated. The total discharge is 

the sum of all station and edge discharge values. 

 

Starting the discharge measurement 

 

Turn the FlowTracker on. Press 

ENTER

 to display the 

Main Menu. Press 

1

 to enter the Setup Parameters Menu. 

 

At the Setup Parameters Menu, review the current settings 

and change the values to meet your requirements. To 
change a displayed value, press the number next to the 

relevant parameter. Press 

ENTER

 to display more menu 

options. For example: Press 

2

 in the Setup Parameters 

Menu to change the Averaging Time. 

The defaults for these parameters are: 

Units: English   

Averaging Time: 40 seconds 

Mode: Discharge 

Salinity: 0.0 ppt (freshwater) 

Discharge Equation:  Mid Section 

 

When you are finished, press 

0

 to return to the Main 

Menu. Now press 

2

 to enter the System Functions Menu. 

 

Field-testing the FlowTracker 

 

Some general tips at the System Functions Menu: 

 

Collect and verify temperature data (press 

4

). 

 

Check battery voltage (press 

5

). 

 

Collect and verify raw data. Ideally, SNR values 

should be >10 dB, but 4 dB is acceptable (press 

6

). 

 

Verify the internal clock is correct (press 

9

). 

 

When you are done, press 

0

 to return to the Main Menu. 

 

Enter site information 

 

Press 

3

 to Start Data Run and display the Data File Name 

Menu. Press 

1

 and enter a file name. To enter text names, 

use the same method as mobile phones (e.g., press 

2

 four 

times for “C”; 

2

A

B

C

). Now press 

9

 to accept the name. 

 

At any time during data collection, press 

8

 (

QC Menu

) to 

enter supplemental data including gauge height, rated 

flow, and user comments. 

 

Collect station data 

 

In the Starting Edge screen, enter the location, depth, 

correction factor, and starting edge using the marked 

buttons on the keypad. Note that LEW/REW stands for 

Left/Right Edge Water. Press 

Next Station

 to continue. 

 

Enter the location, depth, and method of measuring 

velocity (changed by pressing 

/-

). When the 

station information is complete, and the probe is at the 

correct depth and orientation, press the 

Measure

 button. 

 

An updating display will show the measured velocity and 

SNR values. Keep the probe as steady as possible. On 

completion of the averaging time, a summary will be 

displayed. Press 

1

 to accept and go to the next station or 

depth, or press 

2

 to repeat this measurement. These steps 

will be repeated for all stations until 

End Section

 is 

pressed. 

 

Ending the measurement 

 

When 

End Section

 is pressed, the ending-edge 

information is displayed. Enter the information for this 

edge. The 

Previous Station

 and 

Next Station

 buttons can 

also be pressed to review completed stations. Press 

Calc 

Discharge 

to compute the total cross-sectional discharge 

for all completed stations. Press 

0

 to return to the Main 

Menu. 

You must always return to the Main Menu to make 

sure that all data is saved.

 

Отзывы: