Содержание ProDSS

Страница 1: ...st be installed or the bulkhead will be damaged 1 Remove calibration cup see diagram firmly hold the calibration cup with one hand and twist the cup s retaining nut counterclockwise until it loosens d...

Страница 2: ...e battery back up to 80 Expected battery life 20 hours at 25 default LCD brightness 14 hours at 100 LCD brightness Study the main Run screen Confirm the following The battery charge is 10 If not infor...

Страница 3: ...oxes are checked Sample Period s is set to 15 Sample Count is set to 10 c Stabilize Hit the up arrow and select Start Auto Stable You ll return to the Run screen On the Run screen there will be a flas...

Страница 4: ...f your pairs of readings are within the acceptable precision limits resample as needed 9 Record time and full initials of the sampler this should be someone who has been properly trained see Quality A...

Страница 5: ...ination If the port is wet carefully dry with a lint free cloth 2 Apply a thin coat of lubricant to the sensor o rings Wipe off any excess lubricant with a lint free cloth 3 Carefully align the sensor...

Страница 6: ...and let dry before replacing the port cover Battery pack troubleshooting If the instrument is not turning on do the following 1 Attempt to charge the unit Using YSI charging equipment only plug unit i...

Страница 7: ...he logging settings are inaccurate press the system button scroll down to the logging submenu option go to the Continuous Mode and uncheck the box by pressing Enter The box should become unchecked Pre...