background image

 

3.

 

Select your Site

a)

 

Select the 

System 

button 

, then 

highlight 

Logging [single] 

 by using the 

down arrow.  

(Note:  the green band 

indicates which feature is highlighted.)   

Select this feature by hitting Enter. 

b)

 

Before selecting a site, make sure the 

Use Site List

 check box has been 

selected 

.  If not, highlight that row 

and hit Enter. 

c)

 

Highlight 

Site[whatever is in the 

brackets]

.  Hit Enter. 

d)

 

You will see a list of sites.  Scroll down 
until you find your site name, and hit 
Enter.  If you can’t find your site, select 
“test” and note on your field sheet the 
site and time. 

e)

 

The green bar should read 

Select [your 

site].

  Hit Enter, then hit Escape twice.  

This will return you to the Run screen. 

SAMPLING: 

1.

 

Place the probe at the sampling point:

  

Standing downstream or to the side of the 
probe, place it in the main current, in a 
place with steady flow, adequate depth, 
good mixing, and no surface turbulence.  

Do not sample downstream of where 
your team has disturbed the bottom.

  

Sample at mid-depth or 8-12” below the 
surface if the water is deep. You may dangle 
the probe or place it on clean rocks at the 
bottom if the stream is shallow.  

You must 

immerse the probe at least halfway up 
the guard.

  Agitate the probe 

under water to dislodge any 
bubbles at the probe tips. 

2.

 

Stabilize and log readings:

 

Turn on the Auto Stable feature: 

a)

 

Hit the 

Probe button 

, then 

select 

Auto Stable

.

 

 

b)

 

Confirm settings (see image): 

 

Parameters are all set to 
[On] (ISE=pH) 

 

Continuous Mode

 and 

Log 

Samples

 boxes are checked 

 

 

Sample Period(s)

 is set to [15] 

 

Sample Count

 is set to [10]. 

c)

 

Stabilize:

  Hit the up arrow and select 

Start Auto Stable.

  You’ll return to the 

Run screen. On the Run screen, there 

will be a flashing red symbol   next to 
each parameter.  Wait until all of these 
flashing red symbols become green and 

stay green for at least two minutes 

(use the on-screen clock).  Also observe 
the readings (particularly temperature, 
DO, and pH) to look for drift. 

d)

 

Log a reading:

  Once you’re sure that 

all parameters have stabilized, press 
Enter to log a reading. 

The first time 

you hit Enter, it will probably just 
light up the darkened screen; you 
have to hit Enter again before it will 
say “Sample Logged” at the bottom.

  

Anytime the screen is dark, hitting any 
button just brightens it. 

Jumping Values:  If readings won’t stabilize 

…you can still log the readings, but you’ll need 
to estimate values for the non-stable 
parameters.  For instance, if you see turbidity 
values jumping up and down, watch the 
readings over a period of several minutes and 
estimate the average, to the nearest whole 
number, considering both the extremes and the 
most frequent numbers.  In a pinch, we might 
be able to calculate the average at the office. 

3.

 

Record the data:  

If possible, record the 

data directly off the Run Screen right after 
you’ve logged your sample.  It is also 

possible to retrieve the data 
either directly from the meter 
or by linking the meter to a 
computer—see instructions at 
end of protocol.

 

Record the readings on your data 
sheet, in the following order, using 
our rounding convention (see 
Quality Assurance protocol):

 

a)

 

Temperature:

  to nearest 

0.1

°

b)

 

Pressure:

  to nearest 0.01 

inHg 

Содержание ProDSS

Страница 1: ...st be installed or the bulkhead will be damaged 1 Remove calibration cup see diagram firmly hold the calibration cup with one hand and twist the cup s retaining nut counterclockwise until it loosens d...

Страница 2: ...e battery back up to 80 Expected battery life 20 hours at 25 default LCD brightness 14 hours at 100 LCD brightness Study the main Run screen Confirm the following The battery charge is 10 If not infor...

Страница 3: ...oxes are checked Sample Period s is set to 15 Sample Count is set to 10 c Stabilize Hit the up arrow and select Start Auto Stable You ll return to the Run screen On the Run screen there will be a flas...

Страница 4: ...f your pairs of readings are within the acceptable precision limits resample as needed 9 Record time and full initials of the sampler this should be someone who has been properly trained see Quality A...

Страница 5: ...ination If the port is wet carefully dry with a lint free cloth 2 Apply a thin coat of lubricant to the sensor o rings Wipe off any excess lubricant with a lint free cloth 3 Carefully align the sensor...

Страница 6: ...and let dry before replacing the port cover Battery pack troubleshooting If the instrument is not turning on do the following 1 Attempt to charge the unit Using YSI charging equipment only plug unit i...

Страница 7: ...he logging settings are inaccurate press the system button scroll down to the logging submenu option go to the Continuous Mode and uncheck the box by pressing Enter The box should become unchecked Pre...

Страница 8: ...This page left blank for printing...

Отзывы: