pH

 – 

-

 

Do not measure pH in metal containers.

 

-

 

Water currents in streams can cause unstable readings. 

-

 

pH measurements will drift if the sample is not stirred. 

-

 

Do not expose the meter or probe to freezing temperatures. 

-

 

High humidity and cold temperatures may affect meter operation. 

-

 

Be sure that the temperature sensor is immersed in the sample. 

-

 

Buffers degrade from exposure to light - change daily 

SAFETY

 

No special precautions 

PROCEDURES: 

I. 

Configuration – The YSI Pro 1020 instrument is capable of simultaneous measurement of multiple 

water chemistry parameters including temperature, pH, and dissolved oxygen. 

A. 

Instrument overview 

1. 

Meter – The YSI Pro 1020 handheld meter includes internal hardware and 

software for calculation of parameters, a display screen, and a keypad for data 

entry and screen navigation.  The meter includes an internal barometer. 

2. 

Cable – The instrument cable connects the meter to the sensors.  The SLCP uses 

the 6051020 cable with an integrated temperature sensor and two sensor ports. 

3. 

Sensors – two sensors are required to measure all of the water chemistry 

parameters indicated above.  Sensors are installed in the cable bulkhead ports 

indicated below to ensure proper function.  If measurement of one of the 

parameters is not required, the corresponding sensor(s) may be removed and 

replaced with a sensor port plug. 

a. 

temperature  sensor is permanently incorporated into the cable end 

b. 

pH sensor (605101) installed in port 1 

c. 

polarographic dissolved oxygen sensor (605203)  installed in port 2 

II. 

Pro 1020 Key Pad Functions 

Key 

Description 

 

 

Power and Backlight -

 Press once to turn instrument on. Press a second 

time to turn backlight on. Press a third time to turn backlight off. Press and 

hold for 3 seconds to turn instrument off. 

 

Down Arrow - 

Use to navigate through menus, to navigate through box 

options at the bottom of the Run screen and to decrease numerical inputs. 

ENTER 

 

Enter - 

Press to confirm selections, including alpha/numeric key 

selections. 

 

Up Arrow - 

Use to navigate through menus, to navigate through box 

options along the bottom of the Run screen and to increase numerical 

inputs. 

Cal

 

Calibrate - 

Press and hold for 3 seconds to calibrate. Opens Calibrate 

menu from the Run screen.

 

Содержание Pro 1020

Страница 1: ...bes pH 605101 DO 605203 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is achieved if the actual samples to be measured are within 10 C of the calibration solutions Longer cable lengths introduce more error into the readings Dissolved oxygen Chlorine sulfur dioxide nitric oxide and nitrous oxide can affect readings by behaving like o...

Страница 2: ...nsor and two sensor ports 3 Sensors two sensors are required to measure all of the water chemistry parameters indicated above Sensors are installed in the cable bulkhead ports indicated below to ensure proper function If measurement of one of the parameters is not required the corresponding sensor s may be removed and replaced with a sensor port plug a temperature sensor is permanently incorporate...

Страница 3: ...ess the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen 2 Setup Sensors a Press Menu button b Dissolved oxygen sensor 1 Highlight DO Sensor Type Press enter 2 Highlight Polaro black and press enter to confirm 3 Highlight ESC exit press enter 4 Highlight DO Membrane type press enter 5 Highlight 1 25 ...

Страница 4: ...lt in damage e Twist the sensor clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 5 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for ...

Страница 5: ...reading pH mVs in buffer 4 should be 159 to 180 mV from the previous buffer 7 pH mV value pH mVs in buffer 10 should be 159 to 180 mV from the previous buffer 7 pH mV value If the mV readings are outside of these ranges clean the sensor as described in the maintenance section of this protocol 10 Press enter to complete the calibration or press Cal to cancel 11 Calibration Successful will display f...

Страница 6: ...nd across the DO sensor membrane to obtain an accurate reading Be careful not to create a vortex which allows air bubbles to pass against the DO membrane G After at least 5 minutes and once the values stabilize record the measurement IIX Sensor storage A Short term storage The cable assembly is supplied with a sensor storage container or sleeve that attaches to the cable The container is used for ...

Страница 7: ... 15 minutes in clean water Not DI water Soaking in DI water will shorten the life span of the probe containing a few drops of commercial dish washing liquid Gently clean the glass bulb and platinum button by rubbing with a cotton swab soaked in the cleaning solution Rinse the sensor in clean water wipe with a cotton swab saturated with clean water and then re rinse with clean water II Additional c...