IDS WA System

Commissioning

ba76199e03

08/2018

13

3.6.2 Establishing a wireless connection

3.7 Combining several IDS WA-S Charger modules

To be able to charge several IDS WA-S adapters via one USB connection at the 
same time, you can combine up to 5 IDS WA-S Charger modules to form one 
charging station.

1. Open the 

Sensors

 / menu.

automatically (on switching on the meter)

manually (in the measured value display)
– with the softkey 

<F1__>

/[Info] 

– with the softkey 

<F2>

/[

] (MultiLab 4010-3(-2))

2. Select a sensor with 

<

><

>

 

.

3. Open the connection setting for the sensor with 

<ENTER>

.

The IDS WA-S adapter of the selected sensor briefly lights up red.

4. Select the connection setting for the sensor with 

<

><

>

 

 and confirm with 

<ENTER>

.

- - -

Sensor not connected.

LED Sensor 

connected

(LED flashes during data exchange)

 Sensor 

connected

(LED does not flash during data exchange)

5. Use 

<M>

 to quit the 

Sensors

 menu.

The settings are accepted.
The measured value of the sensor is displayed.

Содержание IDS WA

Страница 1: ...OPERATING MANUAL ba76199e03 08 2018 IDS WA System SYSTEM FOR THE WIRELESS OPERATION OF IDS SENSORS IDS WA S IDS WA S Charger IDS WA M OPERATING MANUAL ...

Страница 2: ... System Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email info ysi com Internet www ysi com Copyright 2018 Xylem Inc For the most recent version of the manual please visit www ysi com ...

Страница 3: ...f the IDS WA S Charger 9 3 4 Connecting the IDS WA M adapter to a meter 10 3 5 Connecting the IDS WA S adapter to an IDS sensor 10 3 5 1 Removing the IDS WA S adapter from the plug head of the IDS sensor 11 3 6 Establishing a wireless connection 11 3 6 1 The Sensors menu 11 3 6 2 Establishing a wireless connection 13 3 7 Combining several IDS WA S Charger modules 13 3 8 Mounting an IDS WA S Charge...

Страница 4: ...e battery of the IDS WA S adapter cannot be charged 20 7 4 Status LED of the IDS WA S Charger flashing red green 21 8 Replacement parts and accessories 21 9 Technical data 21 9 1 General features 21 9 2 Adapter IDS WA M 22 9 3 Adapter IDS WA S 23 9 4 IDS WA S Charger 24 Contact Information 27 Ordering Technical Support 27 Service Information 27 ...

Страница 5: ...e IDS WA System IDS WA S Charger To upgrade the IDS WA System equip further IDS sensors with the IDS WA S adapter which is available as an accessory 1 IDS WA M adapter on the meter With an IDS WA M adapter connected the IDS meter registers up to 6 wireless sensors within a radius of up to 10 meters The administering of the sensors is done in the Sensors menu 2 Wireless sensor Unit of IDS WA S adap...

Страница 6: ...nsors and other accessories Safety datasheets of calibration or maintenance accessories such as buffer solutions electrolyte solutions etc 2 2 Safe operation 2 2 1 Authorized use The authorized use of the IDS WA System is exclusively the running of a wireless connection beween IDS sensors with IDS meters Only the operation and running of the analyzer according to the instructions and technical spe...

Страница 7: ...arger to charge the battery in the IDS WA S adapter USB cable to connect the IDS WA S Charger to a USB socket Power pack with USB connection and country specific line adapters Operating manual 3 2 System requirements IDS sensor with plug head connector variant W Available sensors see Internet WA capable meter MultiLab 4010 P 1W IDS MultiLab 4010 P 1 IDS from software version V2 00 MultiLab 4010 2W...

Страница 8: ...rged 3 3 1 Power supply of the IDS WA S Charger 1 Charging slot for the IDS WA S adapter 2 LED indicating the operating condition of the IDS WA S Charger 3 USB B connection for the power supply of the IDS WA S Charger 4 Charging contacts for the IDS WA S adapter 5 Retaining clip for the IDS WA S adapter 1 Plug the USB plug of the USB cable supplied into the USB connector 3 of the IDS WA S Charger ...

Страница 9: ...cts with other USB devices about power resources 1 Open the retaining clip 5 on the IDS WA S Charger 2 Insert the IDS WA S adapter in the IDS WA S Charger The charging contacts of the adapter have to be positioned on the charging contacts 4 of the IDS WA S Charger The function keys of the adapter point upwards 3 Close the retaining clip 5 of the IDS WA S Charger until it clicks into place The stat...

Страница 10: ...nnecting the IDS WA S adapter to an IDS sensor For a working wireless connection the IDS WA S adapter has to be connected to an IDS sensor with plug head connector variant W 1 Turn the IDS WA M adapter so that the notch on the plug of the IDS WA S adapter is aligned to the guidance of the socket 2 Plug the IDS WA M adapter into an IDS sensor socket of the meter 1 Guidance 2 Notch IDS WA M 1 2 ...

Страница 11: ...d sensor 2 Align the guidance 1 of the IDS WA S adapter to the notch 2 in the plug of the plug head sensor and insert the IDS WA S adapter into the unlocked plug of the plug head sensor until is clicks into place 3 Close the locking device on the plug head sensor Make sure that the plug connection is completely dry and clean If necessary clean the plug connection prior to opening 1 Open the lockin...

Страница 12: ...is less than 10 m The radio link is free of visible obstacles 1 Sensor name series number max 8 sensors 2 Setting of the connection to the meter LED Open the connection setting for the sensor with ENTER The IDS WA S adapter of the selected sensor briefly lights up red 3 asterisk The sensor is already connected to another meter If the sensor is connected to another meter the ENTER and CAL keys are ...

Страница 13: ...ter manually in the measured value display with the softkey F1__ Info with the softkey F2 MultiLab 4010 3 2 2 Select a sensor with 3 Open the connection setting for the sensor with ENTER The IDS WA S adapter of the selected sensor briefly lights up red 4 Select the connection setting for the sensor with and confirm with ENTER Sensor not connected LED Sensor connected LED flashes during data exchan...

Страница 14: ...in receptacle for the mechanical positioning of a further IDS WA S Charger module with the pin 4 of the other module 4 Pin for the mechanical positioning of a further IDS WA S Charger module with the socket 3 of the other module 5 Plug for the power supply of a further IDS WA S Charger module via the socket 1 of the other module 1 Plug together up to 5 IDS WA S Charger modules with the aid of the ...

Страница 15: ...nnection the USB connection must provide a current of at least 900 mA 5 x 180 mA 2 Bolt together up to 5 connection rods corresponding to the number of IDS WA S Charger modules 3 Tighten the cap nut with washer on the end of the connection rod with external thread 4 Plug the combined connection rod through the channel 2 up to the stop at the cap nut 5 Tighten a bolt M3x6 with washer onto the end w...

Страница 16: ... is displayed on the meter Establishing the connection for the first time may take several seconds 1 Receptacle for hook Put the IDS WA S Charger on a hook through the receptacle on the underside 2 Bore hole for screwing the module to the wall Fix a screw M5 with head diameter 8 mm to the wall through the bore hole The head of the screw must not protrude from the bore hole so the charging contacts...

Страница 17: ... IDS WA S adapter 1 Function key of the IDS WA S adapter Function key Function Short pressure 1 second The current measured value is stored and output to an interface of the connected meter Long pressure 1 second A measurement with AutoRead is started When the measured value is stable the measured value is auto matically stored and transferred to an interface of the meter that is connected see ope...

Страница 18: ...ndicating can be switched off in the Sensors menu Color Status Meaning GREEN Flashes Data are exchanged between the sensor and meter The LED flashes during each data exchange You can switch off the flashing by selecting the connection status in the Sensors menu RED Flashes once The sensor is identified The function is started for the selected sensor in the Sensors menu with ENTER BLUE Flashes once...

Страница 19: ... Thus avoid contact with acetone or similar detergents that contain solvents Remove any splashes immediately 6 3 Disposal All components of the IDS WA System contain electronics At the end of their operational lifetime the components of the IDS WA System must be returned to the disposal or return system statutory in your country If you have any questions please contact your supplier 7 What to do i...

Страница 20: ...ry out a reset of the meter see operating man ual of your meter Strong damping of the radio signal Remove any obstacles in the radio link e g doors panes etc Run the meter and sensor at a smaller distance Check the signal strength in the menu Sensors F1__ Cause Remedy The maximum number of sensors or adapters is exceeded Disconnect the IDS sensor Disconnect the IDS WA M adapter Cause Remedy Insuff...

Страница 21: ... the IDS WA S adapter in the charging sta tion again Disconnect and then reconnect the charging cable Description Model Order no Adapter for IDS sensors IDS WA S 108141y Adapter for IDS meters IDS WA M 108142y Charging station for IDS WA S adapter IDS WA S Charger 108143y Set for the wireless connection of IDS sen sors with IDS meter IDS WA Kit 108144y Power pack with USB interface NT USB Universa...

Страница 22: ...lable excerpts from the datasheet of the supplier of the BlueTooth LE radio module Guidelines and norms used EMC EC directive 2014 30 EC EN 61000 6 3 EN 61000 3 2 EN 61000 3 3 EN 61000 6 1 FCC Class A Radio data trans mission RED directive 2014 53 EU EN 300 328 EN 301489 1 EN 301489 17 Meter safety EC directive 2014 35 EC EN 60950 IP protection class EN 60529 Dimensions Approx 15 x 18 x 40 mm Weig...

Страница 23: ... IP 66 Test certificates CE FCC Ambient condi tions Storage 25 C 65 C Operation 5 C 55 C Admissible rela tive humidity Yearly mean 75 30 days year 95 Other days 85 Power supply Rechargeable battery Lithium polymer battery 3 7 V 240 mAh Operational life The battery life depends on the power requirement of the sensors connected IDS sensor Battery life h pH ORP Approx 60 h Conductivity Approx 30 h Di...

Страница 24: ...idity Yearly mean 75 30 days year 95 Other days 85 Power supply USB connection and plug connections on the IDS WA S Charger The IDS WA S Charger requires a current of 180 mA to charge the battery of the IDS WA S adapter USB 3 x connections of the PC meet this criterion Power pack with USB connection Helms Man PMB0501200P Input 100 240 V 50 60 Hz 300 mA Output 5 V 1200 mA Primary plugs contained in...

Страница 25: ......

Страница 26: ......

Страница 27: ...g a product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Return form may be downloaded at www ysi com and clicking on the Support tab Telephone 800 897 4151 937 767 7241 Monday through Friday 8 00 AM to 5 00 PM ET Fax 937 767 1058 Email info ysi com Mail YSI ...

Страница 28: ...lyze and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise backed by a legacy of innovation For more information on how Xylem can help you go to www xyle...