OPERATION  

A.  Power ON/OFF 

Press and hold “HOLD – On/Off” 

 key for 3 seconds. The unit will turn on and enter the 

“Measure Mode”.  Press and hold “HOLD – On/Off” key again to turn off.  The instrument has a 
10 minute automatic shutoff. 

 

B.  Set Date and Time  

1. 

Press and Hold “CAL” 

 key until it enters “Date/Time Set Mode”. 

2. 

Press “HOLD” key to select date or time parameters. 

3. Press 

“STORE”   key to increase value or “SCROLL” 

 key to decrease value. 

4. 

Press “CAL” key for 2 seconds to return to “Measure Mode”. 

 

C.  Select Buffer Set 

1. Press and hold “SCROLL” key until unit enters “Buffer Set Select Mode”. 
2. Press “SCROLL” key again to select either 7.00 (7.00/4.01/10.01) or 6.86 (6.86/4.00/9.18) 

buffer sets. 

3. Press “STORE” key for 2 seconds to return to “Measure Mode”.  

Note: 

It should only be necessary to select the buffer set once unless the set of buffers 

changes.    

 
D.  Calibrate pH 

1. Press “CAL” key for 2 seconds and release to enter “Calibration” mode. 
2. Immerse meter into either pH 7.00 or 6.86 buffer solution depending on which Buffer Set you 

are using. ‘WAIT” will flash. 

3. The pH10A meter will determine when the calibration reading is stable. Meter is calibrated 

when “WAIT” disappears.  The unit will automatically enter the next calibration point 5 
seconds later. One/first point calibration is now completed. Press “CAL” key to exit or 
continue to next step for two/second point calibration. 

4. Repeat Steps 2 and 3 for two/second point calibration by immersing the meter in pH 

4.01/4.00 or 10.01/9.18 buffers. 

5. Move to #6 for three/third point calibration or press “CAL” to exit two point calibration. 
6. Repeat Steps 2 and 3 for three/third point calibration. 
7. When “WAIT” disappears unit will automatically return to “Measure Mode”. 

Note: 

For accuracy, it is recommended pH calibration is performed routinely with good buffer 

solutions and a clean electrode or anytime after replacing the electrode. 

 
E.  Measure 

Dip the electrode into the test solution while in the “Measure Mode”. 

 

F.  Hold Data 

1. 

When the pH is stable, press “HOLD” key once to lock the reading on the display. 

2. 

Press “HOLD” key again to unlock reading and the unit will return to “Measure Mode”.  
The unit is now ready for another measurement. 

Note: 

The unit cannot be powered off in “Hold Mode”. 

 
G.  Save Data 

Either in the “Hold Mode” or the “Measure Mode”, press “STORE” key for 2 seconds. “DATA 
SAVE” and “M-XX” will appear indicating the reading has been saved and stored in memory 
location XX.  The unit automatically returns to “Measure Mode”. 

Note: 

The non-volatile memory stores up to 50 sets of data (pH, Temperature, Date, Time) 

even if power is lost. If memory is full, the next set of data will overwrite the first data set stored, 
etc. 

 

Содержание EcoSense pH10A

Страница 1: ...sories and inspect for shipping damages Notify your local dealer or YSI immediately 1 937 767 7241 or environmental ysi com of any damages or missing parts Save packing materials until satisfactory operation is confirmed BEFORE USING A Insert the Batteries 1 Remove the battery cap at the top of the unit 2 Insert the set of batteries included ensuring correct polarities 3 Securely replace battery c...

Страница 2: ...ds later One first point calibration is now completed Press CAL key to exit or continue to next step for two second point calibration 4 Repeat Steps 2 and 3 for two second point calibration by immersing the meter in pH 4 01 4 00 or 10 01 9 18 buffers 5 Move to 6 for three third point calibration or press CAL to exit two point calibration 6 Repeat Steps 2 and 3 for three third point calibration 7 W...

Страница 3: ...ard in this instrument may contain a manganese dioxide lithium coin cell battery that must be in place for continuity of power to memory devices on the board This battery is not user serviceable or replaceable When appropriate an authorized YSI service center will remove this battery and properly dispose of it per service and repair policies ERROR DISPLAYS AND TROUBLESHOOTING pH LCD Display ATC Di...

Страница 4: ...tion prepaid to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement Limitation of Warranty This Warranty does not apply to any YSI product damage or failure caused by i failure...

Страница 5: ... accesorios de su embalaje con cuidado y verifique que no haya sufrido daños durante el envío Avise inmediatamente a su distribuidor local en caso de que haya piezas faltantes o dañadas Guarde los materiales de embalaje hasta que confirme que la unidad funciona satisfactoriamente ANTES DE USAR A Inserte las pilas 1 Quite la tapa de las pilas en la parte superior de la unidad 2 Inserte el juego de ...

Страница 6: ...modo de Calibración 2 Sumerja el medidor en una solución tampón de pH 7 00 ó 6 86 dependiendo de qué juego de solución tampón está usando WAIT Esperar parpadeará 3 El medidor pH10A determinará cuando la lectura de calibración es estable El medidor está calibrado cuando desaparece WAIT La unidad registrará automáticamente el siguiente punto de calibración 5 segundos después Ahora está completa la c...

Страница 7: ...n forma periódica 2 Para el almacenamiento a largo plazo guarde en una solución tampón de pH 4 B Pilas Este instrumento es alimentado por pilas alcalinas que el usuario debe quitar y eliminar cuando las pilas ya no enciendan el instrumento Los requisitos de eliminación varían por país y región y se espera que los usuarios entiendan y cumplan con las exigencias de eliminación de pilas según su loca...

Страница 8: ...ratura Termistor 10 k a 25 C 77º F Información general Duración de la pila Medidor 200 horas Escala de temperatura ambiente 0 0 a 50 0 C 32 a 122º F Pantalla Gráfica de cristal líquido de 98 x 64 Estuche Estuche estanco IP67 Peso 105 g 3 7 onzas REPUESTOS GARANTÍA El medidor pH10A tiene un año de garantía contra defectos de materiales y fabricación contado a partir de la fecha de compra por el usu...

Страница 9: ...manera expresa por YSI ESTA GARANTÍA SE OTORGA EN LUGAR DE CUALQUIER OTRA GARANTÍA EXPLÍCITA O IMPLÍCITA LO QUE INCLUYE TODA GARANTÍA DE COMERCIALIZACIÓN O IDONEIDAD PARA UN PROPÓSITO ESPECÍFICO DE CONFORMIDAD CON ESTA GARANTÍA LA RESPONSABILIDAD DE YSI SE LIMITA A LA REPARACIÓN O REEMPLAZO DEL PRODUCTO LO CUAL SERÁ LA SOLUCIÓN ÚNICA Y EXCLUSIVA QUE TENDRÁ EL COMPRADOR POR CUALQUIER PRODUCTO DEFEC...

Страница 10: ... 2 Premir a tecla HOLD para seleccionar os parâmetros para a data e a hora 3 Premir a tecla STORE para aumentar valores ou a tecla SCROLL para os diminuir 4 Premir a tecla CAL durante 2 segundos para voltar ao Measure Mode C Seleccionar Buffer Set 1 Premir e manter premida a tecla SCROLL até aparecer Buffer Select Mode 2 Premir novamente a tecla SCROLL para seleccionar buffer set de 7 00 7 00 4 01...

Страница 11: ... H Reexibir Dados 1 Em Measure Mode premir a tecla SCROLL durante 3 segundos O conjunto mais recente de dados guardados voltará a ser visualizado 2 Para visualizar dados previamente guardados premir a tecla SCROLL 3 Premir e manter premida a tecla CAL durante 2 segundos para apagar os dados guardados e voltar a Measure Mode 4 OU 5 Premir a tecla STORE durante 2 segundos para regressar a Measure Mo...

Страница 12: ... nova solução tampão Recalibrar ou Substituir eléctrodo ESPECIFICAÇÕES Parâmetros Resolução Exactidão pH 0 00 to 14 00 pH pH 0 01 pH pH 0 02 pH 1 LSD Temp 0 0 to 100 0 C Temp 0 1 C Temp 0 3 C Temp 32 0 to 212 F Temp 0 2 F Temp 0 6 F pH Reconhecimento pH 7 00 4 01 10 01 or pH 6 86 4 00 9 18 Compensação de temp AUTO 0 0 a 100 0 C 32 0 a 212 F Parâmetro de temp tampão 0 0 a 60 0 C 32 0 a 140 F Eléctr...

Страница 13: ...mpo incluído na garantia original ou pelo menos por 90 dias a partir da data da reparação ou substituição Limitação de Garantia Esta Garantia não se aplica a produtos da YSI que tenham sido danificados ou tenham falhado devido a i não serem seguidas as instruções escritas da YSI para instalar operar ou usar o produto ii abuso ou utilização incorrecta do produto iii não fazer a manutenção do produt...

Отзывы: