YSI Incorporated
YSI 550 DO Handheld Instrument
14
keys to increase or decrease the salinity setting. When the correct salinity appears on the LCD
(zero for fresh water), press the ENTER key. The instrument will return to normal operation.
Once the calibration process is complete, the only keys that will remain operational are the MODE
key, the LIGHT key and the ON/OFF key. You can move back and forth from reading dissolved
oxygen in the mg/L mode or the % air saturation mode by pressing the MODE key. If you are
working in a dark area and have difficulty reading the LCD, press and hold the LIGHT key to
activate the back-light of the YSI 550. The ON/OFF key turns the instrument on or off.
For best results:
Each time the YSI 550 DO Instrument is turned off, re-calibrate before taking measurements.
!
Calibrate at a temperature within
±
10°C of the sample temperature.