![Ypsomed YpsoPump Скачать руководство пользователя страница 49](http://html2.mh-extra.com/html/ypsomed/ypsopump/ypsopump_manual_3622935049.webp)
1
3
2
4
49
2/4:
Tap the “Prime cannula” icon.
1/4:
Open the main menu and tap the
“Prime infusion set” icon.
Priming the cannula
After attaching the infusion set to the body, the cannula has to be primed with insulin.
Priming volume for the cannula part, head (volumes are approximate):
5.5 mm steel cannula with base: 0.2 units* (0.002 ml)
8.5 mm steel cannula with base: 0.2 units* (0.002 ml)
6 mm soft cannula with base:
0.3 units* (0.003 ml)
9 mm soft cannula with base:
0.4 units* (0.004 ml)
* 100 U/ml insulin
4/4:
During the priming procedure, the
screen counts the delivered priming
volume upwards until the set value has
been reached. You can cancel the priming
procedure at any time by tapping .
3/4:
The screen for setting the priming
volume appears. For the priming volume,
select a value that is suitable for the
cannula, between 0.1 U and 1.0 U. Confirm
by tapping . The mylife
™
YpsoPump
®
vibrates briefly, and the cannula is primed
with the set amount of insulin.