Warnings
Only use AAA alkaline batteries (LR03) with your YpsoPump
®
. Make sure the battery is not damaged. A damaged battery
could result in a short circuit. If a different type of battery (e.g. a lithium battery or an NiMH rechargeable battery) or a
damaged battery is inserted, there is no guarantee that the insulin pump will be put into operation or operate properly;
the advance warning time up to the “Battery empty” alarm cannot be maintained when a battery is discharged. This can
lead to failure of the preset insulin delivery, which can cause hyperglycemia (read warning in the box on page 5). Make sure
to always carry a replacement battery with you.
Always monitor the self-test attentively and carefully. If you are uncertain whether the self-test has been performed
correctly, repeat the self-test. If errors occur during the self-test (incomplete display, no vibration signal or audible signal),
you must not use the YpsoPump
®
as potential warnings and alarms on the insulin pump can no longer be displayed
and indicated properly. In such a case, contact Customer Service immediately. You find the contact details on the back
of this Quick Start Guide.
Take special care when setting the time of day. If a wrong time of day is set, the timing of the basal rate profile could
be shifted. Only make changes to your settings after consulting your doctor or diabetes advisor. Incorrect programming
of the insulin pump can lead to incorrect dosages of insulin, which can cause hypoglycemia or hyperglycemia
(read warning in the box on page 5).
The instructions for use in this Quick Start Guide are only valid for the following YpsoPump
®
article:
REF 700013408.
You find the reference number of your YpsoPump
®
on your device and on the label of your Starter Kit.