Ypsomed YpsoPump Скачать руководство пользователя страница 32

Stop mode

To switch the YpsoPump

®

 from run 

mode to stop mode, open the main 
menu. Insulin delivery is not possible 
in stop mode.

2/3: 

Confirm by tapping  .  

The YpsoPump

®

 vibrates briefly.

1/3: 

Open the main menu, swipe to 

the left and tap the “Run mode/switch 
to stop mode” icon.

32

Содержание YpsoPump

Страница 1: ...More freedom More confidence With Ypsomed Quick Start Guide Insulin Pump...

Страница 2: ...days expectations in regards to user friendliness As a Swiss company Ypsomed feels committed to a particularly high level of quality This is ensured by a strong reliable handset along with comprehensi...

Страница 3: ...sal rate 17 Programming basal rate profile A 17 6 Inserting an insulin cartridge and priming the infusion set 21 Inserting an insulin cartridge 21 Priming the infusion set 24 Priming the cannula 28 7...

Страница 4: ...rectly repeat the self test If errors occur during the self test incomplete display no vibration signal or audible signal you must not use the YpsoPump as potential warnings and alarms on the insulin...

Страница 5: ...e insulin delivery settings to the changed conditions If you are uncertain contact your doctor or diabetes advisor immediately Check your blood glucose as often as your doctor or diabetes advisor reco...

Страница 6: ...bolus Combination bolus Temporary basal rate function Basal rate profiles A and B Prime tubing Prime cannula Switch basal rate profile Disclaimer The Bluetooth word mark and logos are registered trade...

Страница 7: ...Date Basal rate limit Bolus increment Rotate screen Submenu Main menu Data Settings Stop Run mode Therapy data Alarm history Blind bolus inactive active Bluetooth inactive active Access lock inactive...

Страница 8: ...rtment lid Function button Tubing Cannula soft or steel Tubing cap Adapter Touchscreen AAA alkaline battery LR03 Tape Cartridge compartment with viewing window Infusion set Cannula base Self filled Yp...

Страница 9: ...ill see a picture showing how to insert the battery 2 11 Lock the battery compartment by turning the groove of the battery compartment lid clockwise until it is completely closed using the edge of a c...

Страница 10: ...soon as the battery has been inserted the welcome screen appears and remains until the insulin pump starts the self test The self test is used to check the operability of the YpsoPump and to give the...

Страница 11: ...11 Three test patterns appear on the screen one after the other The 3 3 test rectangles must be dis played in full with uniform brightness 6 11 Then the YpsoPump emits two vibration signals tactile si...

Страница 12: ...e self test by tapping If the self test is not confirmed within 5 minutes after having started up the insulin pump from storage status or after the threaded rod has been returned the insulin pump indi...

Страница 13: ...13 Setting the time of day 9 11 The screen for setting the time of day appears Set the current time of day hours minutes by selecting the desired value Confirm the values set by tapping...

Страница 14: ...n impossible date e g 2017 02 30 the confirma tion icon will be inactive gray and you cannot confirm the date 10 11 After setting the time of day set the current date year month day Confirm your setti...

Страница 15: ...20 seconds and the screen lock is activated If the YpsoPump is not operated in the main menu or in a submenu the screen switches off after two minutes and the screen lock is activated Any changes tha...

Страница 16: ...to the right to return to the locked status screen 6 6 Swipe to the right to access the unlocked status screen no padlock icon 5 6 When the YpsoPump has been unlocked successfully you are in the main...

Страница 17: ...amming two different basal rate profiles A and B If only one basal rate profile is needed profile B has to be programmed in the same way as basal rate profile A 1 8 Open the main menu and tap the Basa...

Страница 18: ...g After each confirmation of an hourly value you automatically move to the next hour The minimum programmable basal rate greater than 0 00U h is 0 05U h The basal rate values between 0 05U h and 1 00U...

Страница 19: ...segment you want to correct Confirm by tapping The YpsoPump vibrates briefly The values are saved and one event per basal rate value is written to the therapy data 6 8 Repeat step 4 8 for each indivi...

Страница 20: ...recommendations from your doctor or diabetes advisor If only one basal rate profile is needed the second one has to be programmed in the same way as the first one By doing so the basal rate will cont...

Страница 21: ...the YpsoPump you may only use a self filled 1 6 ml reservoir YpsoPump Reservoir compatible with the YpsoPump Inserting an insulin cartridge 1 6 Open the main menu and tap the Cartridge change and cur...

Страница 22: ...as described on page 10 Do not insert the new cartridge until the threaded rod has completely returned and the YpsoPump has successfully performed the self test If you insert the cartridge while the t...

Страница 23: ...e cartridge compartment facing upwards Insert a self filled 1 6 ml reservoir 6 6 Place the adapter on the inserted cartridge upright Turn the adapter clockwise to the lock position You should hear a s...

Страница 24: ...is for guidance only It is important that the system is primed free of air and that insulin emerges from the end of the tubing It may happen that more insulin has to be delivered than the priming vol...

Страница 25: ...eck the cartridge for air bubbles If the cartridge contains air bubbles select the appropriate priming volume according to the instructions for use of your YpsoPump Orbit infusion set until there are...

Страница 26: ...dge and the infusion set is primed with the set amount of insulin 5 6 During the priming procedure keep the insulin pump in an upright position with the adapter facing up wards and tap the insulin pum...

Страница 27: ...iming procedure with the appropriate priming volume until there are no more air bubbles in the cartridge in the adapter or in the tubing and until insulin has emerged from the end of the tubing of the...

Страница 28: ...If insulin has emerged from the end of the tubing and there are no more bubbles of air in the cartridge in the adapter or in the tubing attach the infusion set to your body After having attached the i...

Страница 29: ...re at any time by tapping 3 4 The screen for setting the prim ing volume appears For the priming volume select a value that is suitable for your cannula between 0 1 U and 1 0 U according to the instru...

Страница 30: ...m by tapping The YpsoPump vibrates briefly The YpsoPump has two different operating modes Run and Stop The Stop mode and Run mode icons show the current operating sta tus of the insulin pump in each c...

Страница 31: ...3 3 The status screen appears Basal rate delivery starts according to the programmed profile 31...

Страница 32: ...rom run mode to stop mode open the main menu Insulin delivery is not possible in stop mode 2 3 Confirm by tapping The YpsoPump vibrates briefly 1 3 Open the main menu swipe to the left and tap the Run...

Страница 33: ...een The screen of the YpsoPump shows a stop warning if the insulin pump has been in stop mode for more than one hour The stop warning can be triggered prematurely directly after the insulin pump has b...

Страница 34: ...8 Notes 34...

Страница 35: ...35...

Страница 36: ...products and services for people with diabetes This allows users easy discreet and reliable self treatment With Ypsomed Diabetescare self treatment becomes simpler and easier 700014932 10158226 MSTR...

Отзывы: