8.4 Alarm indication and abnormal remedy
Table 8.1
Message
Reason and Remedy
OVER RUN
There is dust on the V-groove... A fiber is
detached from its V-groove. A fiber is dirty.
NO FIBER
The bare fiber is too short... Incorrect
position.
L/R MOTOR
OVER
A motor exceeding the limit position,
causing auto reset.
FAILURE
Arc power is too strong or Arc is unstable.
The gap is too big. Stuff length too small or
no stuff at all
BAD
Splice parameters are inappropriate or arc
power is unstable. Dust on fiber end
profiles. The angle of fiber end profile is
too big.
UNSTABLE
ARC
The interval between two fusion is too
short
. Electrodes need cleaning.
NO ARC
Electrodes are dirty. Worn-out electrodes.
Clean or replace electrodes
L/R -END
BAD
Left or right fiber end profile is too bad;
cleave fiber again.