6.4  COMMAND DETAILS

  COMMAND  

DESCRIPTION

  SET01nn 

Set OUTPUT  MODE

   

01   

Enable voltage output

   

03   

Enable current output

   

04   

Enable RS-232

   

08   

Enable RS-485 output only

   

16   

Enable RS-485 half duplex

   

24   

Enable RS-485 full duplex

  SET02n 

Set  OUTPUT  FORMAT

   

1   

RMYT

   

2   

ASCII

   

3   

ASCII POLLED

   

4   

NMEA

  SET03nn 

Set  BAUD RATE

   

12   

1200

   

48   

4800

   

96   

9600

   

19   

19200

   

38   

38400

  SET04n 

Set  ASCII  WS UNITS

   

1   

MPH 

   

2   

KNOTS

   

3   

KMPH

   

4   

M/S

  SET05c   

ASCII character sensor address (0-9, A-Z, a-z)

  SET06nnn  

Wind speed threshold for polar output (cm/s)

  SET07nnnnn 

Wind speed scale (nnnnn/10000)

  SET08nnnnn 

Direction offset (±nnnnn degrees x 10)

  SET09nn   

Damping factor

  SET10nnnn 

Output interval (0-9999 milliseconds)

  SET11nn   

Direction VOUT (36=0-360, 54=0-540 degrees)

  SET12nnn  

Sample count (3 to 200)

  SET13n   

Wind output format (0/1=Polar/UV)

  SET14n   

Analog error code (1/2/3=None/Lo/Hi)

  SET15n   

ASCII serial resolution(0/1=Lo/Hi)

  SET16n   

Force analog out (0/1/2=Lo/Mid/Hi)

  SET20n   

86004 HTR  (0/1=Enable/Disable)

  SET77 

 

SAVE SETTINGS

 

  XX   

 

Go to OPERATE MODE

  RPT   

 

Report parameter settings

  ??   

 

Command Help list

6.4.1  SET01nn SET OUTPUT MODE

This  enables  and  disables  0-5000  mV  output,  4-20  mA  current 

output,  RS-232,  and  RS-485.  Only  one  of  the  two  analog  output 

types (voltage or current) may be enabled at one time. Both RS-232 

and RS-485 can be enabled at the same time but only one serial 

output may be used.
Add together values shown in 6.4 COMMANDS SET01 to configure 

multiple compatible modes. For example, to enable voltage output 

and RS-232, add the code for each one: 01 + 04 = 05, 

SET0105

. To 

enable only RS-232, 

SET0104

To conserve power, enable only those modes that are needed

6.4.2 SET02n OUTPUT FORMAT

This parameter determines the serial output format.

6.4.3  SET03nn BAUD RATE

Sets the baud rate for RS-232 and RS-485 serial communication. 

Make sure this baud rate is the same as the connected device. 

6.4.4  SET04n ASCII and NMEA WIND SPEED UNITS

Sets wind speed units for polar ASCII, ASCII POLLED, and NMEA 

serial outputs.

6.4.5  SET05c POLL ADDRESS CHARACTER

Sets the sensor address for ASCII POLLED serial format. The default 

is '0' (ASCII 48). This is the address recognized when the 'Ma!' polling 

command is received ('a' is the address character). Valid address 

characters include 0-9, A-Z, and a-z.

6.4.6   SET06nn WIND SPEED THRESHOLD

Sets the wind speed threshold for polar outputs (wind speed and 

direction) to minimize erratic wind direction indications at very low 

wind speeds. This allows the sensor output to mimic a mechanical 

wind vane that retains its orientation when there is no wind.
Wind below the threshold is reported as zero, while the wind direction 

angle is held at the last value when wind speed was above threshold. 

Set threshold in centimeters per second (m/s x 100). The default 

setting is 25 cm/s (0.25 m/s, 0.56 mph).

6.4.7  SET07nnnnn WIND SPEED MULTIPLIER

All wind speed measurements are multiplied by this parameter. The 

default value is 10000 for a multiplier of 1.0000.

6.4.8  SET08nnnnn WIND DIRECTION OFFSET

Use this parameter to add or subtract a wind direction offset.  Value 

is degrees x 10 and may be positive or negative. Wind direction is 

always re-scaled to a 0-360 range after offset is applied. The default 

value is 00000.

6.4.9  SET09nn DAMPING FACTOR

Wind measurement outputs are damped using the following formula:

 

S

damped  

=  [(d-1) * S

damped

 + S

sample

] / d

 

where:

 

S

damped 

= New or last damped wind speed

 

S

sample

  

= New wind speed speed

 

= Damping factor

The default value is 00. This means no damping is applied. High 

damping values at long output intervals can slow the rate at which 

indicated wind values change.

6.4.10  SET10nnnn OUTPUT INTERVAL

Sets  the  time  interval  between  measurements  in  one  millisecond 

increments.  Lower  values  increase  power  consumption  when 

continuous measurements are taken.

6.4.11  SET11nn ANALOG DIRECTION SCALE

Sets wind direction analog output scale to 0-360 or 0-540 degrees. 

Use the 0-540 scale whenever possible to avoid full-scale analog 

output  swings  between  0  and  360  in  variable  north  wind.  (Data 

loggers or display systems may otherwise obtain samples midpoint 

during these transitions causing erroneous readings.)
Logged data in 0-540 form may be re-scaled to 0-360 by subtracting 

360  degrees  from  any  value  greater  than  or  equal  to  360.  The 

default parameter setting is 0-360 for systems that cannot re-scale 

the 0-540 output.

6.4.12  SET12nnn SAMPLE COUNT

The rate at which the sensor internally takes complete sonic wind 

samples  is  greater  than  200  times  per  second.  This  command 

sets the number of internal samples used to calculate the median 

measurement result. 
More internal samples consume more power while providing greater 

immunity  to  conditions  like  turbulent  high-speed  wind.  Fewer 

samples  consume  less  power  while  providing  less  immunity  to 

disruptive conditions.
This command interacts with the OUTPUT INTERVAL setting. More 

samples may require a longer output interval.

Содержание 86004

Страница 1: ...AL INSTRUMENTS INSTRUCTIONS R M YOUNG COMPANY 2801 AERO PARK DRIVE TRAVERSE CITY MICHIGAN 49686 USA TEL 231 946 3980 FAX 231 946 4772 WEB www youngusa com PN 86004 90 REV G092117 ULTRASONIC ANEMOMETER...

Страница 2: ...G Wind Tracker Display 2 5 3 Serial Output Formats and Protocols 2 5 4 Low Power Operation 3 6 0 SETTING SERIAL OUTPUTS AND OPERATING PARAMETERS 6 1 Setup using YOUNG 86SETUP program 3 6 2 Setup using...

Страница 3: ......

Страница 4: ...ady to use Unless otherwise specified the sensor is configured as follows FACTORY DEFAULT CONFIGURATION Analog Voltage Outputs Channel VI1 Wind Speed 0 5000 mV 0 100 m sec Channel VI2 Wind Direction 0...

Страница 5: ...uts since they share connection terminals in the junction box Analog outputs may be configured for either Polar speed and direction or Cartesian UV output format For voltage output with cable lengths...

Страница 6: ...d 1 start 8 data 1 stop bit The sensor must be in COMMAND MODE in order to set parameters Enter COMMAND MODE by sending three ESC characters ASCII 27 in quick succession while the sensor is running Wh...

Страница 7: ...direction to minimize erratic wind direction indications at very low wind speeds This allows the sensor output to mimic a mechanical wind vane that retains its orientation when there is no wind Wind b...

Страница 8: ...sted settings Not all possible setting combinations are shown Using YOUNG 86SETUP program for changing parameters is recommended See wiring diagrams for jumper settings 7 1 FACTORY DEFAULT RS 232 Enab...

Страница 9: ...Page 6 86004 90 G APPENDIX A WIRING CONNECTIONS...

Страница 10: ...Page 7 86004 90 G APPENDIX A WIRING CONNECTIONS...

Страница 11: ...Page 8 86004 90 G APPENDIX A WIRING CONNECTIONS...

Страница 12: ...TH 270 90 180 EXAMPLE WIND AT 45 POLAR WIND SPEED WIND DIRECTION 0 V m s WIND FROM NORTH CARTESIAN U V U m s WIND FROM EAST 86052 OPTIONAL BIRD WIRE ASSEMBLY R M YOUNG COMPANY 2801 Aero Park Drive Tra...

Отзывы: