5-22
IM 01B04K01-02E
No.
Parameter Name
Symbol
Description
Setting Range Factory-set
Value
Display
Conditions
P
Adjustment
Parameters
ADJUST
P03
Zero Adjustment
(Input 1)
ZERO ADJ1
Performs zero adjustment (0% side) on input 1.
n.nnn V RST
n.nnn V INC
n.nnn V HINC
n.nnn V HDEC
n.nnn V DEC
n.nnn indicates the current input value.
Increase or decrease “n.nnn” until the target value is reached.
INC/DEC : Increase/decrease “n.nnn.”
HINC/HDEC : Increase/decrease “n.nnn” more rapidly than INC/
DEC.
RST : When a reset is made, the adjustment values return to their
factory settings.
Displayed
on all
P04
Span Adjustment
(Input 1)
SPAN ADJ1
Performs span adjustment (100% side) on input 1.
The adjustment method is the same as ZERO ADJ1.
P05
Zero Adjustment
(Input 2)
SPAN ADJ2
Zero adjustment (0% side) of input 2
Adjustment method is the same as ZERO ADJ1.
• On SKYD-304 only
P06
Span Adjustment
(Input 2)
SPAN ADJ2
Span adjustment (100% side) of input 2
Adjustment method is the same as ZERO ADJ1.
• On SKYD-304 only
P60
Self Check
SELF CHK
Displays the result (GOOD/ERROR) of the self check.
Displayed
on all
Q
Test Parameters
TEST
Q04
Alarm 1 Forced
Output
RLY1 TEST
Forcibly executes relay output
regardless of the input state. (*1)
DE-
ENERGIZED
ENERGIZED
Displayed
on all
Q05
Alarm 2 Forced
Output
RLY2 TEST
Q60
Self Check
SELF CHK
Displays the result (GOOD/ERROR) of the self check.
*1: After the test ends, press the OK key to cancel the forced output state and set to the normal
operation state.
5.4 Parameter List